background image

Template  version:1 

3/17 

 

SPECIFICATIONS 

1.  Product name: Wrist Blood Pressure Monitor 

2.  Model: KD-723 

3.  Classification: Internally powered, Type  BF applied part,IP22,No AP or APG,Continuous 

operation 

4.  Machine size: Approx. 80mm×60mm×22mm 

5.  Cuff circumference: 14cm ~ 19.5cm(5 1/2

~ 7 11/16

6.  Weight: Approx. 69g (2 7/16oz.) (exclude batteries) 

7.  Measuring method: Oscillometric method, automatic inflation and measurement 

8.  Memory volume: 2

×

60 times with time and date stamp 

9.  Power source: batteries: 2 ×1.5V

SIZE AAA   

10.  Measurement range: 

Cuff pressure: 0-300mmHg 

Systolic:          60-260mmHg 

Diastolic:        40-199mmHg 

Pulse rate:      40-180 beats/minute 

11.  Accuracy:   

Pressure:        ±3mmHg 

Pulse rate:      ±5% 

12.  Environmental temperature for operation: 10

℃~

40

(50

℉~

104

13.  Environmental humidity for operation:   

85%RH 

14.  Environmental temperature for storage and transport: -20

℃~

50

(-4

℉~

122

15.  Environmental humidity for storage and transport: 

85%RH 

16.  Environmental pressure: 80kPa-105kPa 

17.  Battery life: Approx 170 times. 

18.  Wireless Connection: 

Smart Bluetooth 

Frequency Band: 2.400~2.4835GHz 

19. 

All components belonging to the pressure measuring system, including: Pump,Valve, LCD, 

Cuff, Sensor

 

    Note: These specifications are subject to change without notice. 

 
NOTICE 

 

1.  Read all of the information in the operation guide and any other literature in the box before 

operating the unit. 

2.  Stay still, calm and rest for 5 minutes before blood pressure measurement. 

3.  The cuff should be placed at the same level as your heart. 

4.  During measurement, neither speak nor move your body and arm. 

5.  Measuring on same arm for each measurement. 

6.  Please always relax at least 1 or 1.5 minutes between measurements to allow the blood 

circulation  in  your  arm  to  recover.  Prolonged  over-inflation  (cuff  pressure  exceed  300 

mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause 

ecchymoma of your wrist. 

Содержание KD-723

Страница 1: ...OCEDURES 5 1 DOWNLOAD THE FREE APP 5 2 BATTERY LOADING 5 3 CLOCK AND DATE ADJUSTMENT 5 4 CONNECTING THE CUFF TO THE MONITOR 6 5 APPLYING THE CUFF 7 6 BODY POSTURE DURING MEASUREMENT 7 7 TAKING YOUR BL...

Страница 2: ...usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and ac...

Страница 3: ...ored in the memory with date and time stamp The voice function will ease the operation The Electronic Sphygmomanometers corresponds to the below standards IEC 60601 1 2005 corr 1 2006 corr 2 2007 EN 6...

Страница 4: ...0 4 122 15 Environmental humidity for storage and transport 85 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 170 times 18 Wireless Connection Smart Bluetooth Frequency Band 2 400 2...

Страница 5: ...formation regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see part E...

Страница 6: ...f according to local regulations at the end of their usage 3 CLOCK AND DATE ADJUSTMENT a At first the Blood Pressure Monitor is totally off once you insert the battery the Blood Pressure Monitor will...

Страница 7: ...or by pressing START STOP button when the minute is blinking then the time and date is confirmed Note 3 1 The clock format could be set by user 4 CONNECTING THE CUFF TO THE MONITOR The cuff is attache...

Страница 8: ...o not move your arm body or the monitor during measurement Stay still calm for 5 minutes before blood pressure measurement Please keep the cuff clean Clean the cuff by wet soft cloth and mild detergen...

Страница 9: ...ly inflated The blood pressure and pulse will be measured during inflation Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed on the LCD The irregular...

Страница 10: ...Press MEM button again to review the next result See picture 8 6 In this way repeatedly pressing the MEM button displays the respective results measured previously Picture 8 5 Picture 8 6 e When disp...

Страница 11: ...e different levels of blood pressure 12 TECHNICAL ALARM DESCRIPTION The monitor will show HI or Lo as technical alarm on LCD with no delay if the determined blood pressure systolic or diastolic is out...

Страница 12: ...to detect diastolic pressure LCD shows Er 3 Pneumatic system blocked or cuff is too tight during inflation Apply the cuff correctly and try again LCD shows Er 4 Pneumatic system leakage or cuff is to...

Страница 13: ...chnical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measu...

Страница 14: ...ANT INFORMATION REQUIRED BY THE R TTE This product is approved in accordance to R TTE directive transmitter interference This product complies with Industry Canada IC RSS 247 The Bluetooth word mark a...

Страница 15: ...e KD 723 is intended for use in the electromagnetic environment specified below The customer or the user of the KD 723 should assure that it is used in such an environment IMMUNITY test IEC 60601test...

Страница 16: ...NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from struct...

Страница 17: ...aximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz...

Страница 18: ...rgy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particu...

Отзывы: