re-test. 

Speaking,  arm  or  body  movement, 

angry,  excited  or  nervous  during 

testing 

Re-test  when  calm  and  without 

speaking  or  moving  during  the 

test 

Irregular heartbeat (arrhythmia) 

It is inappropriate for people with 

serious arrhythmia to use this 

Electronic Sphygmomanometer. 

 
14. TROUBLESHOOTING (2) 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

LCD shows low battery 

symbol 

 

Low Battery 

Change the batteries 

LCD shows “Er 0” 

Pressure system is unstable 

before measurement 

 

Don’t move and try again. 

LCD shows “Er 1” 

Fail to detect systolic pressure 

LCD shows “Er 2” 

Fail to detect diastolic pressure 

LCD shows “Er 3” 

Pneumatic system blocked or cuff 

is too tight during inflation 

 

Apply the cuff correctly and 

try again 

LCD shows “Er 4” 

Pneumatic system leakage or 

cuff is too loose during inflation 

LCD shows “Er 5” 

Cuff pressure above 300mmHg 

 

Measure again after five 

minutes. If the monitor is 

still abnormal, please 

contact the local distributor 

or the factory. 

LCD shows “Er 6” 

More than 3 minutes with cuff 

pressure above 15 mmHg 

LCD shows “Er 7” 

EEPROM accessing error 

LCD shows “Er 8” 

Device parameter checking error 

LCD shows “Er A” 

Pressure sensor parameter error 

No response when you 

press button or load 

battery. 

Incorrect operation or strong 

electromagnetic interference. 

Take out batteries for five 

minutes, and then reinstall 

all batteries. 

Содержание KD-723

Страница 1: ...OCEDURES 5 1 DOWNLOAD THE FREE APP 5 2 BATTERY LOADING 5 3 CLOCK AND DATE ADJUSTMENT 5 4 CONNECTING THE CUFF TO THE MONITOR 6 5 APPLYING THE CUFF 7 6 BODY POSTURE DURING MEASUREMENT 7 7 TAKING YOUR BL...

Страница 2: ...usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and ac...

Страница 3: ...ored in the memory with date and time stamp The voice function will ease the operation The Electronic Sphygmomanometers corresponds to the below standards IEC 60601 1 2005 corr 1 2006 corr 2 2007 EN 6...

Страница 4: ...0 4 122 15 Environmental humidity for storage and transport 85 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 170 times 18 Wireless Connection Smart Bluetooth Frequency Band 2 400 2...

Страница 5: ...formation regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see part E...

Страница 6: ...f according to local regulations at the end of their usage 3 CLOCK AND DATE ADJUSTMENT a At first the Blood Pressure Monitor is totally off once you insert the battery the Blood Pressure Monitor will...

Страница 7: ...or by pressing START STOP button when the minute is blinking then the time and date is confirmed Note 3 1 The clock format could be set by user 4 CONNECTING THE CUFF TO THE MONITOR The cuff is attache...

Страница 8: ...o not move your arm body or the monitor during measurement Stay still calm for 5 minutes before blood pressure measurement Please keep the cuff clean Clean the cuff by wet soft cloth and mild detergen...

Страница 9: ...ly inflated The blood pressure and pulse will be measured during inflation Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed on the LCD The irregular...

Страница 10: ...Press MEM button again to review the next result See picture 8 6 In this way repeatedly pressing the MEM button displays the respective results measured previously Picture 8 5 Picture 8 6 e When disp...

Страница 11: ...e different levels of blood pressure 12 TECHNICAL ALARM DESCRIPTION The monitor will show HI or Lo as technical alarm on LCD with no delay if the determined blood pressure systolic or diastolic is out...

Страница 12: ...to detect diastolic pressure LCD shows Er 3 Pneumatic system blocked or cuff is too tight during inflation Apply the cuff correctly and try again LCD shows Er 4 Pneumatic system leakage or cuff is to...

Страница 13: ...chnical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measu...

Страница 14: ...ANT INFORMATION REQUIRED BY THE R TTE This product is approved in accordance to R TTE directive transmitter interference This product complies with Industry Canada IC RSS 247 The Bluetooth word mark a...

Страница 15: ...e KD 723 is intended for use in the electromagnetic environment specified below The customer or the user of the KD 723 should assure that it is used in such an environment IMMUNITY test IEC 60601test...

Страница 16: ...NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from struct...

Страница 17: ...aximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz...

Страница 18: ...rgy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particu...

Отзывы: