Restoring the Calibration Data
The zero calibration, span calibration and gravity acceleration values can be
restored to the factory settings values.
Ex. The scale was calibrated with wrong weight, but there is no correct
weight. The values above can be reset to the factory settings.
1. Press the
MODE key twice while
Cal
is displayed to display
Clr
.
2. Press the RE-ZERO key to display
Clrno
.
To switch to
Clrgo
display, press the RE-ZERO key again.
To reset the settings to factory settings, proceed to the next step while
1
Clrgo
is displayed.
3. Press the RE-ZERO key while pressing and holding the MODE key, and
release the MODE key while keeping the RE-ZERO key depressed. The
calibration value is returned to factory settings, and the scale
automatically enters
Cal
display after displaying
end
.
If performing Step 3 while
Clrno
is displayed, the scale will return to
1
Clr
without returning the calibration value to factory settings.
4. Press the ON:OFF key to turn the power off and end the operations.