Span Calibration 

3. Place the calibration weight at the center of the weighing pan and press 

the RE-ZERO key after confirming that the STABLE indicator is lit. 

  Calibration using a weight ends and the scale automatically enters 

weighing mode after displaying   

end

  .  

4. Remove the weight from the weighing pan and press the ON:OFF key to 

turn the power off and end the calibration. 

 

Setting the Gravity Acceleration 

The scale is calibrated with the gravity acceleration value 9.798 m/s

2

 at the 

factory. When the gravity acceleration at your location is different value, you 
should calibrate the scale using a calibration weight. But if the calibration 
weight cannot be prepared, change the gravity acceleration of your scale to 
the value of the area where the scale is used. See “The Value of Gravity 
Acceleration at Various Locations” of this manual. 

 

1. Press the

 

MODE key while   

Cal

      is displayed to display   

9.798 

.

 

2. Press the RE-ZERO key to make the digit to be changed blink. 

Readjust the gravity acceleration value for the scale to be used using the 
following keys. 

MODE            

To change which digit is blinking. 

RE-ZERO 

To increase by +1 the value of the blinking digit. 

3. Press the RE-ZERO key while pressing and holding the MODE key, and 

release the MODE key while keeping the RE-ZERO key depressed. The 
scale automatically enters   

Cal 

    display after displaying   

end

  . 

4. Press the ON:OFF key to turn the power off and end the operations.

 

 
 
 
 
 
 
 
 
 
 

Содержание HT-3000CL

Страница 1: ...1WMPD4003328...

Страница 2: ...ument Avoid installing the scale in direct sunlight excessive dust high humidity high temperature large temperature fluctuations or magnetic fields which may cause problems or malfunctions Drafts and...

Страница 3: ...erated in a commercial environment If this unit is operated in a residential area it may cause some interference and under these circumstances the user would be required to take at his own expense wha...

Страница 4: ...r comparing the weighing value to the upper and lower limit value STABLE indicator Turns on when the weight reading is stable Polarity Shows the polarity of the weighing value NET indicator Turns on w...

Страница 5: ...s may result in excessive force being applied to the load cell and could damage the scale Take care not to apply too much force to the weighing pan Be sure to lock the battery cover again after instal...

Страница 6: ...ify that the reading is zero If not press the RE ZERO key to zero the display When you use a container for weighing place an empty container on the weighing pan and press the RE ZERO key to zero the d...

Страница 7: ...ruple the minimum display from the zero point in grams Nine upper and lower values respectively can be stored Memory number is 1 to 9 Numerical values of the upper and lower limit value are common for...

Страница 8: ...isplaying end the device returns to weighing mode 6 Start comparing using the memory number changed to Confirming the Upper and Lower Limit Value 4 Press the MODE key to display the memory number to b...

Страница 9: ...oggle between plus and minus 8 Press the ENTER key to store the lower limit value The scale then displays the upper limit value after displaying end 9 Press the ENTER key at upper limit value display...

Страница 10: ...00 is an example and may be a different number This is the starting point of the function settings Key Operation MODE To select a function item RE ZERO To change the parameter of each item ON OFF To f...

Страница 11: ...the weighing units enabled or disabled as appropriate There are three types of taels HK general Singapore tael HK jewelry tael and Taiwan tael But one of them can be selected The order of units to be...

Страница 12: ...are not available for use with HT 300CL HK Hong Kong HK tael general is equal to Singapore tael A catty contains 16 taels Newton is a value calculated as follows Newton weight in gram x 9 80665 m s2...

Страница 13: ...function 1 Enables zero trucking function Key lock Misoperation prevention function lock 0 Disables key lock Upper and lower limit value settings and unit switching are available 1 Enables key lock U...

Страница 14: ...lay LED brightness Adjusts LED brightness for the comparator display Cpi 0 Dark 1 2 Bright Decimal point Selects the symbol used for the decimal point pnt 0 Dot 1 Comma Initialization init Resets sett...

Страница 15: ...ayed When calibrating using a weight select a calibration weight from the following The weight must be purchased separately by the customer HT 300CL 300 g 0 01 g HT 3000CL 3000 g 0 1 g HT 500CL 500 g...

Страница 16: ...ibration weight But if the calibration weight cannot be prepared change the gravity acceleration of your scale to the value of the area where the scale is used See The Value of Gravity Acceleration at...

Страница 17: ...ch to Clrgo display press the RE ZERO key again To reset the settings to factory settings proceed to the next step while 1Clrgo is displayed 3 Press the RE ZERO key while pressing and holding the MODE...

Страница 18: ...pply 4 x AA batteries R6P LR6 or AC adapter Battery life Approx 450 hours When using at 23 C with comparator disabled and using alkaline batteries Approx 220 hours When using at 23 C with comparator a...

Страница 19: ...acity 1550 0 15500 2550 0 25500 ct Minimum display 0 5 5 0 5 5 Weighing capacity 3 00 30 0 5 00 50 0 N Minimum display 0 01 0 01 0 01 0 01 The Value of Gravity Acceleration at Various Locations Amster...

Страница 20: ...4 1235 550485 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 6 33 817 150 749 817 Manhattan Bldg 33 Gukjeg...

Отзывы: