AD-4406 Weighing Indicator 

 

18. Specifications 

Analog Input and A/D Conversion 

Input sensitivity 

Up to 0.2 µV/division 

Input signal range 

-1 mV to 15 mV 

Load cell excitation voltage 

5V DC  5%, 60 mA with sense voltage input 

Load cell drive capacity 

Maximum 4 x 350 

Ω

 

load cells 

Temperature Zero 

(0.2 µV + 0.0008 % of zero adjustment voltage)/  (typ.) 

  coefficient 

Span 

0.0008%/   of reading (typ.) 

Non-linearity 0.01 

% of full scale 

Maximum input noise 

Less than 0.4 µVp-p 

Input impedance 

10 M

Ω

 or more 

A/D conversion method 

Integrating dual slope type 

A/D resolution count 

40000 counts 

A/D conversion rate   

Approximately 10 times/s (unstable), 5 times/s(stable) 

Maximum display resolution  20000 (permissible 40000) 

Digital Section 

Measurement display 

7 segment, Liquid crystal display 

 

Character height 

25 mm 

State indicator symbol 

Minus sign, Zero point, Stable, Net, Comparison result 
Storing accumulated data, Percentage, Low battery 
Various state indicators (triangle 1, 2, 3) 

Unit 

kg, g, t (lb, oz, lb-oz / catty, tl, catty-tl : depends on the 
region) 

Interface 

RS-232C interface 

Serial interface for communication (D-Sub 9P male contact) 

RS-422/485 interface 

Serial interface for communication, control (terminal block)   

Current Loop output 

20 mA, Passive type (DIN 8) 

Analog output 

4-20 mA, free scaling output (terminal block) 

External Control Input 

3 (or 1) contact input terminals, selected functions (DIN 8) 
3 point (DIN 8 or terminal block) 
Capacity: 50V AC/DC, maximum current 100 mA (resistive load)
Comparison mode selection 
HiHi, Hi, OK, Lo, LoLo output for limit comparison 

Relay output 

Zero band, preliminary, free fall, final for setpoint comparison 

General 

Power supply 

7 to 10VDC (AC adapter, battery (6 x C size) 

Power consumption 

Less than 10VA 

Operation temperature 

-10  to +40  (14  to 104 ) 

Operation humidity 

85% R.H. (no condensation) 

Mass 

650 g approximately 

Dimensions 

192 (W) x 96 (H) x 86 (D) mm 

Accessories 

Refer to "4.4. Accessories" 

 

 

18. Specifications 

Содержание AD-4406 OP-04

Страница 1: ...INSTRUCTION MANUAL 1WMPD4000506B AD 4406 OP 03 04 05 07 08 11 WEIGHING INDICATOR INSTRUCTION MANUAL...

Страница 2: ...or the user to operate the instrument This is a hazard alert mark 2009 A D Company Limited All rights reserved No part of this publication may be reproduced transmitted transcribed or translated into...

Страница 3: ...4 4 Accessories 9 5 Calibration 10 5 1 Items of the Calibration Mode 10 5 2 Calibration Procedure 11 5 2 1 Configuring a Weighing Instrument 11 5 2 2 Specifying the Range and Unit 12 5 2 3 Specifying...

Страница 4: ...ts 39 11 2 2 Simple Batch 40 11 2 3 Setting the Parameters of Setpoint Comparison 41 12 Hold Function 42 12 1 Setting the Hold Functions 42 13 RS 232C Interface OP 04 OP 05 OP 08 44 13 1 Specification...

Страница 5: ...ial environment If this unit is operated in a residential area it may cause some interference and under these circumstances the user would be required to take at his own expense whatever measures are...

Страница 6: ...niversal Flexi Coms function to customize the protocol of outputting data using the serial interface The following interfaces are either standard or are available as an option The RS 232C serial inter...

Страница 7: ...e humidity is less than 45 Plastic and isolators are apt to be charged with static electricity 3 2 Power Supply When an AC adapter is used A stable power source must be used an unstable power source w...

Страница 8: ...ells The load cell drive is 5VDC 5 between EXC and EXC the maximum current 60 mA 3 4 Adjustment of the Load Cell Output Caution Use a metal film resistor in the range of 50kohm to 500kohm with a good...

Страница 9: ...voltage E 5000 mV Weighing interval D 0 05 kg Weighing capacity 300 kg 2 0 1 750 05 0 3 5000 Therefore regard the instrument as a good design 3 6 Installing an Option Board Caution Do not remove any...

Страница 10: ...of measure Unit used to weigh ZERO key The key to zero the current display 11 key The key to select the digit while in the setting mode TARE key The key to perform tare 12 key The key to select a valu...

Страница 11: ...Standby display Zero error when turning the display on If the ESC key is pressed the current weight value is displayed Overload display Remove any load from the load cell immediately It may cause dam...

Страница 12: ...r to function as a weighing instrument Set values do not need to be changed again unless the indicator itself is replaced For details refer to 5 2 1 Configuring a Weighing Instrument CAL 0 Calibrates...

Страница 13: ...ot necessary to set the gravity acceleration correction when calibrating the scale with a calibration mass at the place where the scale is used Enter the stable weight data while the STABLE mark is di...

Страница 14: ...tion first unit is displayed and the alternate unit second unit is blinking ENTER The key to store the data displayed and proceed to the next step ESC The key to proceed to the next step without chang...

Страница 15: ...tion greater than the first range Press the ENTER key to store the parameter and proceed to the next step When pressing the ESC key regardless of what is displayed the indicator will proceed to the ne...

Страница 16: ...press the ENTER key while the STABLE mark is off detecting motion Arrange the condition using the F00 filter function Digital Input Step10 Pressing the key a stored input voltage parameter of the zer...

Страница 17: ...he key a stored input voltage parameter of the span is displayed in the unit of mV V Adjust the input voltage using the following keys key To select a digit key To select a value ENTER key To store th...

Страница 18: ...ng this function is not required Example 1 The gross display Specified parameters First range Range 50 00 kg division 0 02 kg Second range Range 100 00 kg maximum capacity division 0 1 kg Display 0 kg...

Страница 19: ...setting Up to three weighing points can be specified Caution This function does not improve repeatability or hysteresis Use the mass on the condition that Lnr 1 Lnr 2 Lnr 3 Do not press the ENTER key...

Страница 20: ...ld the ON OFF key and press the ESC key to display CAnCEL Then no parameters are changed and the calibration mode is finished Press the ON OFF key to turn the display off Step 2 The parameter is displ...

Страница 21: ...9 809 m s2 Chicago 9 803 m s2 Rio de Janeiro 9 788 m s2 Copenhagen 9 815 m s2 Rome 9 803 m s2 Cyprus 9 797 m s2 San Francisco 9 800 m s2 Djakarta 9 781 m s2 Singapore 9 781 m s2 Frankfurt 9 810 m s2 S...

Страница 22: ...e maximum capacity Reduce calibration mass or digital input Err 5 The selected calibration mass is smaller than the minimum division Err 6 The new input sensitivity is less than 0 2 V division Increas...

Страница 23: ...e pressing the key 2 When in the weighing mode press both the F1 and F2 keys at the same time When you are in the function setting F00 will be displayed Operating Item Step 1 Select an item using the...

Страница 24: ...ro drift may be not detected If CF00 is 1 then 0 1 6 or 7 can be selected only for F01 If CF00 is 1 the initial setting for F01 is 7 0 No motion detection 1 0 5 d 0 5s 2 1 0 d 0 5s 3 2 0 d 0 5s 4 3 0...

Страница 25: ...Serial data output 2 x 3 Accumulation M x 4 totAL display x 5 Setting setpoint x 6 Start comparator batch x 7 Stop comparator batch x 8 Stop buzzer sound x 9 Unit change F10 Type2 F1 key function x A...

Страница 26: ...ory code reading when shorting the terminal 19 Start averaging to hold F13 EXT1 Function selection of external input 20 Release holding data 0 17 19 and 20 Same as F 13 F14 EXT2 18 Memory code BCD 1 0...

Страница 27: ...1 All time 0 Start stop 1 Initial settings are 11 21 31 40 10 11 Tare when start 1 20 21 Stop after Full by key 0 Auto 1 30 31 Over Under judgment after dribble flow 0 Batch finish 1 F23 F22 10 12 Typ...

Страница 28: ...Accumulated value x 8 Accumulation count x 9 Stable Over x A Displayed weight x B Gross weight x C Net weight x D Tare weight x E Weight type G N T refer to CF06 F34 Serial data format 1 Initial value...

Страница 29: ...eter Description 0 Stream mode command is not acceptable 1 Manual mode command is effective 2 Auto print mode command is effective 3 Auto print mode command is effective 4 When accumulation automatica...

Страница 30: ...mode Description of Manual Print Object F37 1 F40 1 Operation When pressing the key set to output data the stable weight data is output just once Description of Auto Print Mode Object F37 2 F37 3 F40...

Страница 31: ...t TARE ZERO in motion TARE at negative gross 0 Not to execute Not to execute 1 Execute Not to execute 2 Not to execute Execute CF04 TARE ZERO in motion TARE at negative gross value 3 Execute Execute 0...

Страница 32: ...e sound of the comparison buzzer Unit for changing the unit HoLd for starting the average or releasing the hold data The function set as first priority will operated when the F1 or F2 key is released...

Страница 33: ...he power off the tare data is cleared CF04 0 1 Weighing Tare Operation Place the tare on the weighing unit Press the TARE key to store the tare weight after the STABLE mark turns on The display change...

Страница 34: ...umulated data can select result of comparison is OK only or all result Accumulation Condition F21 of the F function In case of manual accumulation mode press the key to accumulate weight data when the...

Страница 35: ...ction to display totAL and the accumulated data Step 2 Press and hold the key for more than 3 seconds The display blinks once and the data accumulated before accumulating the last weight data is displ...

Страница 36: ...setting 2 Automatic Fixed format 3 Manual 4 Automatic Format 1 selected at F34 5 Manual Accumulated data output at accumulated data display 6 Automatic Format 2 selected at F35 Step 1 Press the key a...

Страница 37: ...a is lost when power is turned off The active code memory number can be changed by key switch external control input or a command via the serial interface Changing the Code Memory Number by External I...

Страница 38: ...tputs the result to the three relays of HI OK and LO Use this comparison when judging whether a weight is proper Set the F function F22 to 1 2 or 3 to use the upper lower limit comparison and F22 to 4...

Страница 39: ...tored in non volatile memory and are not lost even if power is turned off or the batteries are depleted When the displayed value becomes an overload positive over HI over is output When the displayed...

Страница 40: ...tting Order and Display for the Weight Check Mode F22 MODE Display 1 2 3 4 5 1 Upper Setpoint Upper Lower Lower Class Hi Lo 1 Comparator HI LO 2 Upper Setpoint Target Upper Lower Lower Class tG Hi Lo...

Страница 41: ...ut can be changed to OVER or UNDER by setting F24 The triangle 2 is not displayed OVER UNDER is judged always and the result is output to the display and the relay F22 8 Check weighing 2 Output and Co...

Страница 42: ...Free fall are off from weighing completion to the next start 5 Weighing completion can be output by setting F24 and F25 6 The start signal can be accepted after weighing completion even if the weight...

Страница 43: ...F26 of the F Functions and the value does not belong to a specific code memory but is used commonly Setting Order and Display for Setpoint Comparison and Simple batch F22 MODE Display 1 2 3 4 5 7 Set...

Страница 44: ...weight display is blinking during the averaging period The output data in the averaging period is the actual weight at the time The weighing unit is blinking when the weight display is in the hold st...

Страница 45: ...averaging Continue Stop Stop Weight is going to over at averaging Pending Stop Pending Hold key input at averaging Stop Stop Stop Release input from external input at averaging Stop Stop Stop Release...

Страница 46: ...on parity for 8 data bits Start bit 1 bit Stop bit 1 bit Code ASCII Terminator CR LF CR CR 0Dh LF 0Ah Connector D sub 9 pin connector Circuit and Pin Connection Pin No Signal name Direction Descriptio...

Страница 47: ...re replaced by spaces except for the decimal point Unit F34 35 F Unit In case of CF07 0 the unit length is 2 digits In case of CF07 1 the unit length is 3 digits Depending on the circumstances an A D...

Страница 48: ...UFC commands 13 3 1 Commands to Request Data Request Display Data 1 When receiving this command returns the displayed data immediately Template RW Command Reply Request Display Data 2 When receiving...

Страница 49: ...mat of 9 2 Display and Operation Output of Accumulated Data Is Zero When receiving this command returns at zero point or no immediately Template RZ Command Reply When at ZERO Not at ZERO 13 3 2 Comman...

Страница 50: ...Template CA Command Reply Changing the Weight Unit Changes the weight unit Template UC Command Reply Changing the Code Memory Changes the Code Memory number Template SC m m code memory number 0 4 Com...

Страница 51: ...ue m code memory number 0 4 n setpoint order number refer to Setting the setpoint Command Reply Set Zero Band Sets the F26 value zero band of the comparison The decimal point is not necessary Template...

Страница 52: ...zing the print out of the printer or efficient data collection Output data is not only the indicator s data status but also the characters at will It can output the control code of the printer depends...

Страница 53: ...ple DL10 1 0 second delay strings Output the specified strings enclosed by single quotation itself is described using three single quotations Example A D Set data bits 8 bits if using the 8 bit charac...

Страница 54: ...B Input Receive B terminal 5 TRM Terminator resistance 100 RS 422 RS 485 6 SG Signal ground 7 HI Output Relay output HI 8 OK Output Relay output OK 9 LO Output Relay output LO Relay output 10 COM Rela...

Страница 55: ...RS 485 14 RS 422 485 Relay Output OP 03 RDA RDB SDA SDB SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM SDA SDB RDA RDB TRM A B Some computers have terminators inside S...

Страница 56: ...p a signal width of more than 100 ms for the On time and Off time ON Make More than 100 ms More than 100 ms OFF Break The relays output the result of comparison Solid state relay Maximum voltage 50VDC...

Страница 57: ...30 of the F Functions to 1 and set F31 F32 and F33 Item Parameter Remarks F30 0 No output Initial setting Data output 1 Analog output Must be set to 1 2 Serial in out 1 Data output 3 Serial in out 2 F...

Страница 58: ...signal width of more than 100 ms for the On time and Off time ON Make More than 100 ms More than 100 ms OFF Break The relays output the result of comparison Solid state relay Maximum voltage 50VDC Ma...

Страница 59: ...Each output terminal can be connected to either the plus or minus inputs of the peripheral unit Data Format The data format is the same as that of the initial setting of F34 of RS 232C data format The...

Страница 60: ...ison result Storing accumulated data Percentage Low battery Various state indicators triangle 1 2 3 Unit kg g t lb oz lb oz catty tl catty tl depends on the region Interface RS 232C interface Serial i...

Страница 61: ...AD 4406 Weighing Indicator Page 59 18 1 Dimensions 18 1 Dimensions...