AD-4406 Weighing Indicator
F 2
ENTER
F 2
ENTER
Stored input voltage
New input voltage
mV/V
+/-
ZERO
TARE
mV/V
Place nothing on
the weighing unit
(No load)
Turn on STABLE mark
Zero calibration
To Span Calibration
5.2.8. Zero
Calibration
Step 8 Check the CAL 0 display.
Select a zero calibration method to adjust the zero
point.
Weighing input
(Normal way)
The adjustment method
with nothing on the
weighing unit.
To step 9
Digital input
The numerical way to enter
a load cell output voltage.
To step 10
Weighing Input
Step 9 Place nothing on the weighing unit. Press the ENTER
key after the STABLE mark has turned on. The new
zero point parameter will be stored. Proceed to step11.
ESC key ......... To proceed to step 11 without changing
the zero point parameter.
Caution Do not press the
ENTER
key while the STABLE
mark is off (detecting motion). Arrange the
condition using the F00 filter function.
Digital Input
Step10 Pressing
the
+/-
key, a stored input voltage parameter
of the zero point is displayed in the unit of mV/V. Adjust
the input voltage using the following keys.
key .......... To select a digit.
key .......... To select a value.
ENTER key .... To store the zero point parameter and
proceed to the next step.
ESC key ......... To proceed to the next step without
changing the parameter.
5.2.9. Span
Calibration
Step11 Check the capacity display after CAL F is displayed for 2 seconds.
Select a span calibration method to adjust the capacity.
Weighing a mass less than
the maximum capacity
The method to weigh a mass less
than the maximum capacity.
To step 12
Weighing a maximum
capacity mass
The method to weigh a mass
equivalent to the maximum capacity.
To step 14
Digital input
The numerical way to enter a load cell
output voltage.
To step 16