![AND AD-4406 OP-04 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/and/ad-4406-op-04/ad-4406-op-04_instruction-manual_2941308028.webp)
Page
26
AD-4406 Weighing Indicator
Data Output
Item
Parameter
Description
* 0
No data output
1
Analog output
Set F31, 32, 33
2 Serial
output
F30
Data output
3
Serial output (Zero suppressing)
RS-232C,
RS-422/485
*: Initial settings (In case of installing RS-232C, F30 is set to 2. )
Analog Output
Item
Parameter
Description
* 0
Displayed data
1 Gross
data
F31
Output data
2 Net
data
F32
Weight value at 4 mA output
-999999 to 999999
(Initial setting is 0)
F33
Weight value at 20 mA output
-999999 to 999999
(Initial setting is 10000)
Polarity is changed
by pressing the +/-
key.
*: Initial settings
Serial Data Format
Item
Parameter
Description
x 0
Terminator
x 1
Device ID (selected at F06)
x 2
Code number
x 3
Data number
x 4
Result of comparison
x 5
(Reserved)
x 6
(Reserved)
x 7
Accumulated value
x 8
Accumulation count
x 9
Stable/Over
x A
Displayed weight
x B
Gross weight
x C
Net weight
x D
Tare weight
x E
Weight type (G /N / T, refer to
CF06)
F34
Serial data format 1
Initial value
19
」
,2E
」
,3A,4F,
E
50
F35
Serial data format 2
Initial value
17,2F,
E
30
x F
Weight unit (refer to CF07)
100,10:order of
output (1-999)
Select with the
1
1
key.
1: output data
Select with the
key.
"
E
xxx" indicates
the end of data to
output.
Press
NET/GROSS to
expand or shorten
the output data length.
1 “
」
” (right hand of the figure) of the display indicates that the comma (F46=0, semi-colon
for F46=1) is output after the data. To turn this mark on and off, press the ZERO and
TARE keys at the same time.
Output of initial settings of F34 (19
」
,
2E
」
, 3A, 4F, 50) is like “ST,GR,12345.6kg”.
2 The data number (parameter 3) increments automatically with each serial data output
from 1 to 99999 (the next of 99999 is 1). The starting number can be set by the keys
(F2 + NET/GROSS).
6.2. F-Functions