Maintenance and care
qTOWER
3
84 / qTOWER
3
84 G
Edition 02.19
33
To remove reagent residue fill the affected wells with max. 20 µL distilled water
or ethanol. Wait for approx. 1 min before you suck out the liquid again. Repeat
the process until the background signal is in the normal range.
8.3
Disinfecting the device
WARNING! Biological hazard!
Clean the qTOWER
3
84 with particular care after analyzing potentially infectious material.
Wear suitable protective equipment, e.g. protective gloves.
NOTICE
The only suitable cleaning method for the housing is wipe disinfection.
When using spray disinfectants there is a risk that the liquid enters the sensitive electronic
system through the air vents. If the disinfectant has a spray nozzle, apply disinfectant to a
suitable cloth before using it on the device.
Avoid contamination by handling sample substances with care.
Wipe spilled samples or reagents immediately with an absorbent cloth or piece
of paper.
If the qTOWER³ 84 is used for the analysis of infectious material, great care
must be taken because the qTOWER³ 84 cannot be decontaminated as a whole
device.
Immediately remove visible contamination by suitable means. Do not allow sol-
vents to enter the device.
The sample block is also suitable for wipe and spray disinfection The only suit-
able cleaning method for the housing is wipe disinfection.
Device part
recommended disinfect-
ants
Provider
Sample
chamber
Descosept AF
Meliseptol HBV (cloths)
Dr. Schuhmacher GmbH
B. Braun company
Housing
Descosept Spezial
Dr. Schuhmacher GmbH
Pay attention to the efficacy spectrum of the listed disinfectants with regard to your cus-
tomer-specific decontamination requirements!
Only the use of ethanol or isopropanol containing disinfection agents is ap-
proved for wipe disinfection. The use of disinfection agents with ingredients or
concentrations that differ from the recommended disinfection agents invalidates
all warranty claims regarding damage to the device or non-effectiveness.
If the qTOWER
3
84 has to be sent to Analytik Jena AG for service, a prior de-
contamination must be carried out and documented (see "Return" p. 41).