Technical data
qTOWER
3
84 / qTOWER
3
84 G
Edition 02.19
17
3
Technical data
Block format
384 wells
Sample volume
10 – 20 µL
Lid temperature
30 – 110 °C
Temperature gradient
25 °C (
qTOWER
3
84 G only
)
Max. heating rate*
max. 4 °C/s,
Ø
3.8 °C/s
(depending on consumables used)
Max. cooling rate*
max. 2 °C/s,
Ø
1.7 °C/s
(depending on consumables used)
Heating rate adjustment
min. 0.1 °C/s
Temperature uniformity
(15 s after starting the clock)
±0.15 °C at 55 °C
±0.25 °C at 72 °C
±0.50 °C at 95 °C
Temperature range
3 °C – 99 °C
Control accuracy
± 0.1 °C
Temperature increments
min. 0.1 °C/cycle
Time increments
min. 1 s/cycle
Heated lid
Manual opening mechanism, automatic contact
pressure
Heated lid contact force
comply with 30 kg, automated
Dimensions
(H x W x D)
59 cm x 27.5 cm x 27.5 cm
70 cm x 27.5 cm x 50 cm when opened
Mass
30 kg
Noise level
45 dB
Energy supply
Operating voltage
Line frequency
100 V, 115 V, 230 V
50 – 60 Hz
Power consumption (max.)
max. 850 W
Instrument fuses
2 x 10 AT / 250 V
Safety circuits
Sensory monitoring of the closure and locking of
the sample chamber,
over-temperature protection in the heated lid
Operating conditions
Air pressure: 0.7 bar to 1.06 bar
Max. permissible altitude: 2000 m
Humidity: max. 70 %
Operating temperature: +15 °C to +35 °C