Alarm
Description
Screen with Operator Instructions
9.4.2 Too Long In
Jacket Charge
Occurs if jacket does
not reach set
temperature within
allotted time.
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions
9.4.3 Too Long To
Seal Door
Occurs if door seal
does not reach 5 psig
within allotted time.
STATUS . . ALARM!
231
TOO LONG IN JACKET CHARGE
CHAMBER: 00.0 F 0.0 psig
JACKET: 00.0 F
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL JACKET
TEMP. IS REACHED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
-> IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
232
TOO LONG IN JACKET CHARGE
-> JACKET DID NOT REACH REQUIRED
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. STEAM PRESSURE LESS THEN 50 PSIG
-> CHECK STEAM SUPPLY PIPING
2. STEAM REGULATOR MALFUNCTION
-> REPAIR
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S09
4. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
FEED
EXIT
STATUS . . ALARM!
227
TOO LONG IN SEAL DOOR
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
-> IF CLOSED, OPEN VALVE AND
RESELECT CYCLE
3. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Содержание C Series
Страница 62: ...6 18 P129394 176 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 146: ...10 2 P129394 176 Operator Manual Service Procedures Figure 10 1 Serviceable Components ...
Страница 153: ...10 9 Service Procedures Operator Manual P129394 176 Figure 10 1 Piping Schematic ...