8-4
P129394-176
Operator Manual
Routine Maintenance
8.3 Weekly
Maintenance
Procedure: Flush
Chamber Drain
Flush chamber drain as follows whenever the line becomes clogged:
1.
Turn off steam supply valve. Wait until jacket pressure is zero.
Wait until chamber has cooled to room temperature.
2.
Remove chamber drain strainer (
Figure
). Clean strainer
using procedures given
, if necessary.
3.
Pour a solution of 60 mL (1/4 cup) of Liqui-Jet 2 Instrument
Detergent (
Contact your local STERIS representative
), and 500 mL
(1 pint) of hot water into the drain. Solution may puddle in the
bottom of the chamber.
Should the detergents in
Step 3
be unavailable, you may use a
hot solution of 15mL (1 tablespoon) of trisodium phosphate to
500 mL (1 pint) of hot water.
4.
Open door and place strainer back in drain.
WARNING – BURN HAZARD:
Allow sterilizer and
accessories to cool to room
temperature before
performing any cleaning or
maintenance procedures.
!