9-6
P129394-176
Operator Manual
Troubleshooting
Alarm
Description
Screen with Operator Instructions
9.2.6 Over Sterilize
Temperature
NOTE:
The control
temperature is the value the
sterilizer maintains during the
sterilize phase. This value is
2°F above the cycle sterilize
temperature set point. The
over temperature value is
added to the control
temperature value to arrive at
the temperature which causes
this cycle alarm.
Occurs if chamber
temperature exceeds
the maximum sterilize
temperature (Control
tempe
over-temperature
value).
Screen with Service Instructions
STATUS . . ALARM!
235
OVER STERILIZE TEMPERATURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY CONTINUE CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
SERVICE
HELP
ABORT
SILENCE
ALARM
STATUS . . SERVICE INFORMATION:
236
OVER STERILIZE TEMPERATURE
-> STERILIZE TEMPERATURE IS ABOVE
SETPOINT BY MORE THAN PRESCRIBED AMOUNT
CAUSES AND CORRECTIONS:
1. STEAM PRESSURE MORE THAN 30 PSIG
-> CHECK STEAM SUPPLY PIPING
2. CHAMBER STEAM TRAP MALFUNCTION
-> REPAIR
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S09
-> REPAIR S02
4. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
MORE
HELP
STATUS
PAPER
FEED
EXIT