9-2
P129394-176
Operator Manual
Troubleshooting
Full Print Format Shown
Figure 9-2. Typical Alarm Printout
9.1.2 Typical Alarm Printout
When an alarm occurs the printer automatically generates a printout,
typically listing alarm name, time alarm occurred, current chamber
status and any associated sensor temperature. See
Figure
.
9.2 In-Cycle Alarms
The following alarm screens appear only during cycle operation:
* ALARM
##/##/##
PRESSURE IN CHAMBER
F 10:07:23A
61.7C
2.34P
Alarm
Description
Screen with Operator Instructions
9.2.1 Too Long In
Charge
Occurs if chamber
does not reach the set
temperature within the
allotted time.
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions
STATUS . . ALARM!
205
TOO LONG IN CHARGE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPETE CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK STEAM SUPPLY VALVE
-> IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
206
TOO LONG IN CHARGE
-> CHAMBER DID NOT REACH STERILIZE
TEMPERATURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. STEAM PRESSURE LESS THAN 50 PSIG
-> CHECK STEAM SUPPLY PIPING
2. STEAM REGULATOR MALFUNCTION
-> REPAIR
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S09
-> REPAIR S02
4. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
EXIT
ABORT
SERVICE
MODE