![Amsco C Series Скачать руководство пользователя страница 125](http://html1.mh-extra.com/html/amsco/c-series/c-series_operators-manual_2936189125.webp)
9-5
Troubleshooting
Operator Manual
P129394-176
Alarm
Description
Screen with
Service
Instructions
Too Long In Air Break
(Continued)
Screen with Operator Instructions
9.2.5 Under Sterilize
Temperature
Occurs if chamber
temperature drops
below sterilize
temperature.
Screen with Service Instructions
STATUS . . SERVICE INFORMATION:
226
TOO LONG IN AIR BREAK
-> CHAMBER DID NOT AIR BREAK
VACUUM TO 1 PSIG WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. AIR INLET FILTER PLUGGED
-> REPAIR
2. SOLENOID VALVE MALFUNCTION
-> REPAIR S01
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
EXIT
PAPER
FEED
STATUS . . ALARM!
223
UNDER STERILIZE TEMPERATURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY RESTART STERILIZE TIME
AND CONTINUE CYCLE AFTER SET
TEMP. IS REACHED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
224
UNDER STERILIZE TEMPERATURE
-> CHAMBER TEMPERATURE DROPPED BELOW
STERILIZE TEMPERATURE BY UNDERTEMP
TEMPERATURE VALUE
CAUSES AND CORRECTIONS:
1. STEAM PRESSURE LESS THAN 50 PSIG
-> CHECK STEAM SUPPLY PIPING
2. CHAMBER STEAM TRAP MALFUNCTION
-> REPAIR
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S09
-> REPAIR S02
4. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
EXIT
PAPER
FEED