SECTION D: OPERATION, FUNCTION, AND USE 

 

D-1 General Use 

The AMPTEC 620LK is a milliohmmeter designed to safely measure low level DC 

resistance from 0.01 to 1999.9 m

Ω

. The 620LK meter uses internal solid state voltage 

reference(s) that permit the product to be used immediately after the unit has been 

turned on (temperature range limits permitting). If you flip the carrying case lid 

completely open, and connect the carrying case strap to the unit’s front buckle, the unit 

can be carried hands free. 

 

To use the 620LK meter to make a resistance measurement, first connect the desired 

measurement leads to the Trident connector on the front panel of the meter. Place the 

measurement leads across the unknown resistance value that needs to be measured. 

At this point, the 620LK meter should start updating the LCD screen once an electrical 

connection has been made. After approximately 3 seconds, the meter will produce a 

beeping sound, and the backlight will turn on to illuminate the LCD screen. If the 

measurement leads are removed from the component that was being measured, the 

meter will hold the resistance reading on the screen. The 620LK meter will resume 

measuring resistance once the measurement leads sense a new applied resistance 

value, or by pressing the “Auto Measure Reset” button. 

 

If you have any other general use questions, please contact AMPTEC. 

 

D-2 Front Panel 

The below sub-sections, D-2.1 through D-2.7, provide information on the features, 

design, and uses of the 620LK meter’s front panel. The below photo also shows the 

details of the 620LK’s front panel. 

 

 

 

 

 

 

 

 

 

 

 

D-2.1 Test Leads and Chassis Ground Connections 

The AMPTEC 620LK meter test lead connection uses a 4 terminal-in connector on its 

front panel, as seen in the diagram on the next page. 

 

The unit’s front panel chassis ground panel jack (which also accepts an external banana 

jack) is wired to its all metal case, metal front panel, rear panel, battery compartment,  

Содержание 620LK

Страница 1: ...13231 Rooster Springs Rd Austin TX 78737 T 512 858 4045 E info amptec com amptec com AMPTEC RESEARCH MODEL 620LK OPERATION AND MAINTENANCE MANUAL...

Страница 2: ...ck Feature Information p 9 D 3 620LK Padded Carrying Case Photo and Information p 11 D 4 Detailed Use Description p 11 SECTION E MAINTENANCE REPAIR AND CALIBRATION E 1 General Maintenance Repair Calib...

Страница 3: ...s UAVs drones missiles etc Continuity determinations on fuses electrical cable connections heavy duty metal to metal connections Bonding measurements for metal to metal connections Featuring an all me...

Страница 4: ...replacement see section E 9 1 on page 17 4 Perform a visual test lead integrity check It is recommended that the user perform a basic visual inspection of their test lead set to confirm there is no d...

Страница 5: ...play is 20 000 counts 4 digit liquid crystal display with a back light that indicates autolock See section D 2 7 on page 9 for additional information about the autolock feature Low Battery Indication...

Страница 6: ...a jack test lead set 630 304G 4 4 Wire banana jack test lead set with ground 630 305 4 Twin single banana jack lead set 630 401 4 Single point handheld Kelvin probes 630 405K 10 heavy duty spiked Kelv...

Страница 7: ...d start updating the LCD screen once an electrical connection has been made After approximately 3 seconds the meter will produce a beeping sound and the backlight will turn on to illuminate the LCD sc...

Страница 8: ...wires then terminate in a variety of configurations depending upon the accessory The 620LK keyed access connector also makes it virtually impossible during normal operation to incorrectly connect lea...

Страница 9: ...quid crystal display and high quality the display itself should have a life lasting many years Should the display become damaged or stop operating properly it can be replaced at an AMPTEC facility The...

Страница 10: ...ther to produce the final autolocked measurement which is the reading displayed while the backlight is on To assist the user the meter produces an audible beeping indicator along with the lit backligh...

Страница 11: ...tem D AUTO MEASURE RESET BUTTON Press and hold to reset a measurement that has been locked Item E Power On Off Switch with LED Press and hold for 2 seconds to power the unit on The display backlight a...

Страница 12: ...s the test probes against a conductor meaning continuity is sensed When ready to make another measurement simply place the probes back on the test items and the meter will automatically make another m...

Страница 13: ...duct lines and is capable of fixing nearly any issue your instrument may have AMPTEC maintains one sole facility for both the manufacturing and repair of its products AMPTEC RESEARCH CORP 14101 Hwy 29...

Страница 14: ...00 mA DC level of test current On a 1 resistor the 620LK 100mA output should get a 100mV drop 0 10V 0 10A For DVM scaling on the 2 range voltage drop is amplified using a x10 for range scaling hence a...

Страница 15: ...2 Check for a reading of 0 000 3 Put the output leads across a 0 100 resistance standard 4 The display should read 100 00 m 10 counts depending on the resistor s accuracy If the above test proves unt...

Страница 16: ...ated display of the measurement The meter will continue to sample and hold and then resample as long as it senses any continuity Finally connect the meter via the 630 304 test leads across a precision...

Страница 17: ...he main PCB back in and replace the front panel and screws E 9 Battery Compartment Design and Description The metal battery compartment is powder coated to make it chemically resistant to battery leak...

Страница 18: ...This certification was gained in 2013 and is maintained with quarterly inspections by an independent party AMPTEC also follows UL 913 s quality standards plant wide including the adoption of a qualit...

Страница 19: ...____________________________________________________________ ____________________________________________________________ ____________________________________________________________ _________________...

Отзывы: