background image

620LK Operation and Maintenance Manual – Rev. J – Updated January 2020 

Page 13

   

 

 

 

 

 

 

 

 

 

 

 

 

13231 Rooster Springs Rd. 

Austin, TX 78737 

T:

 (512) 858-4045 

E:

 [email protected] 

amptec.com 

SECTION E: MAINTENANCE, REPAIR, AND CALIBRATION 

 

E-1 General Maintenance, Repair, and Calibration Information 

The AMPTEC 620LK meter was designed with the intent of using high quality, strong, 

and reliable components both internally and externally. As a result, the unit is extremely 

reliable and rarely requires repair. Should the unit require maintenance or repair that 

cannot be performed by the end user, we recommend sending the product to AMPTEC 

RESEARCH’s facility to perform this service. Malfunctions can sometimes be the result 

of bad test lead/connection wiring, wrong connections, misinterpretation of 

specifications, low battery levels, or a misunderstanding of how to use the product. 

 

The below sections will provide information on sending your unit to AMPTEC for repair, 

as well as how the user can perform some maintenance, troubleshooting, and 

calibration techniques on their own. 

 

E-2 Returning the 620LK to AMPTEC for Maintenance, Repair, and Calibration 

AMPTEC RESEARCH has been proudly providing product support for all of our 

equipment for the past 20 years. Part of this support is offering maintenance, repair, and 

calibrations for our instrumentation. Our staff is highly trained across our product lines 

and is capable of fixing nearly any issue your instrument may have. AMPTEC maintains 

one sole facility for both the manufacturing and repair of its products: 

 

   

 

 

   

AMPTEC RESEARCH CORP. 

 

   

14101 Hwy 290. W. 

 

   

Bldg. 2000A 

 

   

Austin, TX, USA 78737 

 

Should you have a product that you would like to be sent in for maintenance, repair, or 

calibration, it can be sent to the address above.

 Before sending in a product to our 

facility, you must contact a member of our staff and receive an RMA #.

 We can be 

reached via email at 

[email protected]

 or phone at 512-858-4045. The purpose 

process of this is so that we can obtain your full contact information and properly 

coordinate the return of the AMPTEC meter. 

 

Once we have received your meter, we will diagnose the issue for free and provide you 

with pricing information. If diagnosing the problem requires one of our technicians to 

open the product in order to reach the internal circuitry, a calibration fee will be required 

for us to send the product back to you. This is because we have a plant-wide policy that 

all instrumentation must be calibrated before leaving our facility – it would be unethical 

to allow out of ordnance products to be sent to an end user. 

 

E-3 Circuit Board Operation Overview 

There is one internal printed circuit board (PCB) that performs the digital ohmmeter 

resistance measurement. The AMPTEC 620LK meter uses a constant current source 

circuit and DVM circuitry all on one PCB. When the constant current source is combined 

with the digital voltmeter circuit, the result is a stable DC resistance measurement.  

Содержание 620LK

Страница 1: ...13231 Rooster Springs Rd Austin TX 78737 T 512 858 4045 E info amptec com amptec com AMPTEC RESEARCH MODEL 620LK OPERATION AND MAINTENANCE MANUAL...

Страница 2: ...ck Feature Information p 9 D 3 620LK Padded Carrying Case Photo and Information p 11 D 4 Detailed Use Description p 11 SECTION E MAINTENANCE REPAIR AND CALIBRATION E 1 General Maintenance Repair Calib...

Страница 3: ...s UAVs drones missiles etc Continuity determinations on fuses electrical cable connections heavy duty metal to metal connections Bonding measurements for metal to metal connections Featuring an all me...

Страница 4: ...replacement see section E 9 1 on page 17 4 Perform a visual test lead integrity check It is recommended that the user perform a basic visual inspection of their test lead set to confirm there is no d...

Страница 5: ...play is 20 000 counts 4 digit liquid crystal display with a back light that indicates autolock See section D 2 7 on page 9 for additional information about the autolock feature Low Battery Indication...

Страница 6: ...a jack test lead set 630 304G 4 4 Wire banana jack test lead set with ground 630 305 4 Twin single banana jack lead set 630 401 4 Single point handheld Kelvin probes 630 405K 10 heavy duty spiked Kelv...

Страница 7: ...d start updating the LCD screen once an electrical connection has been made After approximately 3 seconds the meter will produce a beeping sound and the backlight will turn on to illuminate the LCD sc...

Страница 8: ...wires then terminate in a variety of configurations depending upon the accessory The 620LK keyed access connector also makes it virtually impossible during normal operation to incorrectly connect lea...

Страница 9: ...quid crystal display and high quality the display itself should have a life lasting many years Should the display become damaged or stop operating properly it can be replaced at an AMPTEC facility The...

Страница 10: ...ther to produce the final autolocked measurement which is the reading displayed while the backlight is on To assist the user the meter produces an audible beeping indicator along with the lit backligh...

Страница 11: ...tem D AUTO MEASURE RESET BUTTON Press and hold to reset a measurement that has been locked Item E Power On Off Switch with LED Press and hold for 2 seconds to power the unit on The display backlight a...

Страница 12: ...s the test probes against a conductor meaning continuity is sensed When ready to make another measurement simply place the probes back on the test items and the meter will automatically make another m...

Страница 13: ...duct lines and is capable of fixing nearly any issue your instrument may have AMPTEC maintains one sole facility for both the manufacturing and repair of its products AMPTEC RESEARCH CORP 14101 Hwy 29...

Страница 14: ...00 mA DC level of test current On a 1 resistor the 620LK 100mA output should get a 100mV drop 0 10V 0 10A For DVM scaling on the 2 range voltage drop is amplified using a x10 for range scaling hence a...

Страница 15: ...2 Check for a reading of 0 000 3 Put the output leads across a 0 100 resistance standard 4 The display should read 100 00 m 10 counts depending on the resistor s accuracy If the above test proves unt...

Страница 16: ...ated display of the measurement The meter will continue to sample and hold and then resample as long as it senses any continuity Finally connect the meter via the 630 304 test leads across a precision...

Страница 17: ...he main PCB back in and replace the front panel and screws E 9 Battery Compartment Design and Description The metal battery compartment is powder coated to make it chemically resistant to battery leak...

Страница 18: ...This certification was gained in 2013 and is maintained with quarterly inspections by an independent party AMPTEC also follows UL 913 s quality standards plant wide including the adoption of a qualit...

Страница 19: ...____________________________________________________________ ____________________________________________________________ ____________________________________________________________ _________________...

Отзывы: