background image

one corner of the DVM PCB has a nylon washer

acting as an insulator/separator from its corner

stand off support)
must be

. Save this nylon washer as it

be returned to its PCB metal standoff

location when the adjustment procedure is fin-

ished. Slightly lift and tilt the right hand side of
the DVM PCB to give you access to all the

DMM trimpot locations ( needed for voltage
and current measurement adjustments). All

voltage adjustments should be performed with

DC Voltage Standard input
with an AC Voltage Standard. DC Voltage input

from a DCV standard provides the most stable

AMPTEC 601ES voltage reading for calibration
purposes.

Select the 600V range and input 0.00 VDC

from the DC Voltage standard. Adjust RV101 to

read 000.0 volts.

Still on the 600V range input 600.0 VDC

from the DC Voltage standard. Adjust RV102 to

read 600.0 volts.

Select the 200V range and input 100.00 VDC

from the DC Voltage standard. Adjust RV105 to

read 100.0 volts.

Select the 20V range and input 10.000 VDC

from the DC Voltage standard. Adjust RV104 to

read 10.000 volts.

Select the 2 V range and input 1.0000 VDC

from the DC Voltage standard. Adjust RV103 to

read 1.000 volts.

You may now verify all the above voltage

ranges with input from an AC Voltage standard
to determine compliance.

Select the 600V range and provide 600VAC

@60Hz input, then verify the range is in specifi-

cation.

Select the 200V range and provide 100VAC

@60Hz input, then verify the range is in speci-
fication.

Select the 20V range and provide 10VAC

@60Hz input, then verify the range is in specifi-

cation.

Select the 2 V range and provide 1.0 VAC

@60Hz input, then verify the range is in speci-
fication.

All trimpot adjustments ( i.e RV401)

described for this section are also located on the

DMM PCB.

then later verify

3.

4.

5.

6.

7.

8.

A)

B)

C)

D)

F-3-6 AC/DC Current Adjustments

1)

the battery compartment

). When adjusting

RV405,

which is indicated as

a negative sign.

Select the 20 Kohm range. Connect the Kel-

vin clips to the 10 Kohm standard resistor.

Select the 2 0 Ohm range. Connect Kelvin

leads to the AMPTEC 601ES Safety DMM.

Short the Kelvin Clip leads at the end

.

do not over adjust

F-3-3. 2.0 Ohm Range Adjustments con't

(zero)

Adjust RV402 on the DVM PCB for a

reading of 1.0000 ohms. Over adjustment is

indicated by a negative polarity sign.

F-3-2. Full Scale Adjustment
1.

F-3-3. 2.0 Ohm Range Adjustments
1.

.

There are no adjustments

necessary for the 20 Ohm, 200 Ohm, and 20
Kohm ranges.

2.

2.

3.

Adjust RV401 for a display indication of

10.00 Kohms.

for a dis-

play indication of .0000 ohms.

Connect the test leads to a 1.0 Ohm resistance

standard.

Check the 20 KOhm range with the

10.0 KOhm Standard Resistor.

Adjust RV403 on the DVM PCB

F-3-4. Verification of Each Ranges Accuracy

F-3-5 AC/DC Voltage Adjustments

Check the 20 ohm range with a 10 ohm resis-
tance standard. Check the 200 ohm range with a

100 ohm resistance standard. Check the 2.0

KOhm range with the 1.0 Kohm Standard
Resistor.

All resistance

ranges must be within the specifications out-

lined in Chapter B.

1. The adjustments for this section is difficult to
access, we recommend a simple

for all AC/DC Voltage and AC/DC

current measurements to

determine if the

AMPTEC 601 Safety DMM is adequately in

specification. If the AMPTEC 601ES DMM is

"in spec" you do not need to perform the adjust-
ment procedure described in the steps below. All

the adjustments described in this section are
performed on the DMM PCB (

).

2. Keeping all ribbon cables and wiring intact

between the various PCBs,

remove

the DVM PCB from its standoff mounts. (

verification

check first

first

carefully

see figure H-14

for PCB location, the DMM PCB is essentially

located below the DVM PCB and the I source
PCB

note:

20

Содержание 601ES

Страница 1: ...RESEARCH AMPTEC CALIBRATED MODEL SER NO DATE DUE Tamper ProofSeal Sticker Calibrationvoidifremoved 601ES 601ES 503 05APR2005 05APR2006 I press hold 2 0 KOHM 20 OHM 200 OHM 2 0 OHM 20 KOHM FUNCTION RA...

Страница 2: ...cifications C 1 Available Accessories and Options D 1 General D 2 Front Panel Features and Operation D 3 Complete 601ES Explosive Safety DMM Front Panel Diagram D 4 4 Wire Resistance Measurement 601ES...

Страница 3: ...Procedure F 3 1 Zero Offset Adjustment F 3 2 Full Scale Adjustment F 3 3 2 0 Ohm Range Adjustments F 3 4 Verification of Each Ranges Accuracy F 3 5 AC DC Voltage Adjustments F 3 6 AC DC Current Adjust...

Страница 4: ...osive Safety Igniter Circuit Test industry must be held down for 2 seconds The AMPTEC 601ES DMM Safety Meter has a circuit that indicates when it is time To replace the batteries They can be replaced...

Страница 5: ...ince Nickel Metal Hydrides are re chargeable they offer a battery life cycle a hundred times greater than any alkaline Be sure to charge the Nickel Metal Hydride batteries when taken out of their orig...

Страница 6: ...around the battery holder and secure Cut excess plastic Next wrap the black and red wire around the battery holder and place the holder into the compartment Lastly replace lid and screw in screws to...

Страница 7: ...ange Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range 2 0 Amperes DC or AC RMS range Accuracy 0 50 of range 2 0 Ohm Range thru 20 KOhm ranges...

Страница 8: ...e point of the probe tip connection eliminating 98 of the cable resistance Gold plated tips and banana jack ends provide an optimal low thermal emf connection gold plated single points gold plated gol...

Страница 9: ...banana plugs at one end The other end is terminated with an alligator clip red and black These alligator clips will measure 2 wire method properly with either upper or lower jaw connected to the condu...

Страница 10: ...once the access screws have been removed for calibration These handles are for carrying the 601ES DMM around when the lid is removed They are strictly for calibration access he main Power switch to tu...

Страница 11: ...M is factory set using internal PCB jumpers to automatically turn off after 30 minutes of inactivity no range or function change to conserve battery power 5 Automatic Turn Off Protection Fuse Stray Cu...

Страница 12: ...g the zero knob past the stops will severely damage the precision of the instrument and violate the instruments warranty 2 20 and 200 ohm ranges 10 Case Chassis Ground Jack not tied to earth ground A...

Страница 13: ...s go in continuous decades from 2 0 Ohms to 20 Kilo Ohms The selection of the 2 0 Ohm lowest resistance range is accomplished by rotating the range knob Align indicator mark on knob with the desired r...

Страница 14: ...ck to the AMPTEC 601ES DMM through the jack Depending upon the magnitude of the resistance offset some of the offset may be corrected zero ed out for in the AMPTEC 601ES DMM by adjusting the zero knob...

Страница 15: ...g Circuitry fatigued adapter wire plug wire splice loose the fewer the connections the fewer the problems The LCD display will indicate when it is time to replace the batteries LO BAT LO BAT The The A...

Страница 16: ...atsoever If you turn on display does not come on Explosive Safety DMM m AMPTEC Explosive Safety DMM and the If the AMPTEC 601ES Explosive Safety DMM exhibits problems that cannot be eliminated by revi...

Страница 17: ...only when the integrator is de integrating At the end of a conversion the data remaining in the results counter is latched decoded and multiplexed to the liquid crystal display The analog section blo...

Страница 18: ...n also turn the unit off the same way after a certain time period has elapsed This is to save battery life See FIG H 1 The micro can turn on TR201 TR209 using IC201 and I 202 These fets are used to se...

Страница 19: ...y Out of Specification Section B No adjustments are needed Verify each function and range with the appropriate calibration standards to determine if the AMPTEC 601ES DMM meets its published accuracy s...

Страница 20: ...PCB then later verify 3 4 5 6 7 8 A B C D F 3 6 AC DC Current Adjustments 1 the battery compartment When adjusting RV405 which is indicated as a negative sign Select the 20 Kohm range Connect the Kel...

Страница 21: ...press to measure button and Adjust RV502 and repeat this step 5 dual button press operation until the AMPTEC 601ES reads 0050 A 5 milliamperes Return both the and the s to their standoff PCB mounts Re...

Отзывы: