F-3-6 AC/DC Current Adjustments con't

(should be outputting ~50 mV at this

point)

2) Select the 2.0 amp range and provide 1.0

ampere DC from the DC Current calibrator.
Keep in mind, you have to press/hold the stray
current pushbutton to measure AC/DC current.
Adjust

(located in the middle of the

DMM PCB) to read 1.000 ampere. Verify AC
current with 1.0 amp@ 60Hz input from an
AC/DC Current Calibrator to read within spec
(see section B for specifications)

Since the voltage and current measurements
have just be calibrated in the previous steps, a
verification check of the

and

signal

source is

to any signal

source adjustments (only adjust when neces-
sary). Again, keep all ribbon cables and wiring
intact between the PCBs,

remove the

from its standoff mounts. Slightly

lift and tilt the left hand side of the I-Source PCB
to give you access to the

trimpots

below it ( needed only for access to the signal
source adjustments).

Select the 2.0 V range.

Press and hold the 50mV/5mA Signal Source

pushbutton.

Adjust

located on the

(underneath the I-Source PCB) to read .0500 V .

4.) Next select the 2 amp range.
5.) Press/hold the 50 mV/5mA Signal Source
button at the same time as the 2.0 stray current
press to measure button and

. Adjust RV502 and repeat this step 5

dual button press operation until the AMPTEC
601ES reads .0050 A ( 5 milliamperes).

Return both the

and the

s to their standoff PCB mounts. Return the

nylon washer/insulator to the DVM PCB corner
standoff mount. With all ribbon cabling intact,
re-secure both PCBs to their standoff mounts.
Carefully replace the top panel and umbilical

RV401

F-3-7 50mV/5mA Signal Source Adjustments

50mV

5 mA

recommended prior

I-Source PCB

DMM PCB

1.)

2.)

3.)

RV501

DMM PCB

note the display

reading

I-Source PCB

DVM

PCB

carefully

cabling to it's bezel mount and re-secure the

calibration access screws located around the
perimeter of the 601ES top plate.

To replace low batteries first unscrew the
battery compartment lid and lift. Remove the
battery holder, plastic tie and low batteries.

Replace with a new set of (4) “AA” batteries
according to polarity. Using the spare plastic

tie included in the battery compartment, wrap

the plastic around the battey holder and secure.
Cut excess plastic. Next, wrap the black and

red wire around the battery holder and place the

holder into the compartment. Lastly replace lid
and screw in screws to secure. See diagram

above. *Please note: After replacing battery
holder with (4) new ‘AA’ batteries; secure with

a new plastic tie lock.

F-4. Battery Replacement Instructions
The battery pack replacement process is:

21

Содержание 601ES

Страница 1: ...RESEARCH AMPTEC CALIBRATED MODEL SER NO DATE DUE Tamper ProofSeal Sticker Calibrationvoidifremoved 601ES 601ES 503 05APR2005 05APR2006 I press hold 2 0 KOHM 20 OHM 200 OHM 2 0 OHM 20 KOHM FUNCTION RA...

Страница 2: ...cifications C 1 Available Accessories and Options D 1 General D 2 Front Panel Features and Operation D 3 Complete 601ES Explosive Safety DMM Front Panel Diagram D 4 4 Wire Resistance Measurement 601ES...

Страница 3: ...Procedure F 3 1 Zero Offset Adjustment F 3 2 Full Scale Adjustment F 3 3 2 0 Ohm Range Adjustments F 3 4 Verification of Each Ranges Accuracy F 3 5 AC DC Voltage Adjustments F 3 6 AC DC Current Adjust...

Страница 4: ...osive Safety Igniter Circuit Test industry must be held down for 2 seconds The AMPTEC 601ES DMM Safety Meter has a circuit that indicates when it is time To replace the batteries They can be replaced...

Страница 5: ...ince Nickel Metal Hydrides are re chargeable they offer a battery life cycle a hundred times greater than any alkaline Be sure to charge the Nickel Metal Hydride batteries when taken out of their orig...

Страница 6: ...around the battery holder and secure Cut excess plastic Next wrap the black and red wire around the battery holder and place the holder into the compartment Lastly replace lid and screw in screws to...

Страница 7: ...ange Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range 2 0 Amperes DC or AC RMS range Accuracy 0 50 of range 2 0 Ohm Range thru 20 KOhm ranges...

Страница 8: ...e point of the probe tip connection eliminating 98 of the cable resistance Gold plated tips and banana jack ends provide an optimal low thermal emf connection gold plated single points gold plated gol...

Страница 9: ...banana plugs at one end The other end is terminated with an alligator clip red and black These alligator clips will measure 2 wire method properly with either upper or lower jaw connected to the condu...

Страница 10: ...once the access screws have been removed for calibration These handles are for carrying the 601ES DMM around when the lid is removed They are strictly for calibration access he main Power switch to tu...

Страница 11: ...M is factory set using internal PCB jumpers to automatically turn off after 30 minutes of inactivity no range or function change to conserve battery power 5 Automatic Turn Off Protection Fuse Stray Cu...

Страница 12: ...g the zero knob past the stops will severely damage the precision of the instrument and violate the instruments warranty 2 20 and 200 ohm ranges 10 Case Chassis Ground Jack not tied to earth ground A...

Страница 13: ...s go in continuous decades from 2 0 Ohms to 20 Kilo Ohms The selection of the 2 0 Ohm lowest resistance range is accomplished by rotating the range knob Align indicator mark on knob with the desired r...

Страница 14: ...ck to the AMPTEC 601ES DMM through the jack Depending upon the magnitude of the resistance offset some of the offset may be corrected zero ed out for in the AMPTEC 601ES DMM by adjusting the zero knob...

Страница 15: ...g Circuitry fatigued adapter wire plug wire splice loose the fewer the connections the fewer the problems The LCD display will indicate when it is time to replace the batteries LO BAT LO BAT The The A...

Страница 16: ...atsoever If you turn on display does not come on Explosive Safety DMM m AMPTEC Explosive Safety DMM and the If the AMPTEC 601ES Explosive Safety DMM exhibits problems that cannot be eliminated by revi...

Страница 17: ...only when the integrator is de integrating At the end of a conversion the data remaining in the results counter is latched decoded and multiplexed to the liquid crystal display The analog section blo...

Страница 18: ...n also turn the unit off the same way after a certain time period has elapsed This is to save battery life See FIG H 1 The micro can turn on TR201 TR209 using IC201 and I 202 These fets are used to se...

Страница 19: ...y Out of Specification Section B No adjustments are needed Verify each function and range with the appropriate calibration standards to determine if the AMPTEC 601ES DMM meets its published accuracy s...

Страница 20: ...PCB then later verify 3 4 5 6 7 8 A B C D F 3 6 AC DC Current Adjustments 1 the battery compartment When adjusting RV405 which is indicated as a negative sign Select the 20 Kohm range Connect the Kel...

Страница 21: ...press to measure button and Adjust RV502 and repeat this step 5 dual button press operation until the AMPTEC 601ES reads 0050 A 5 milliamperes Return both the and the s to their standoff PCB mounts Re...

Отзывы: