CHAPTER F - CALIBRATION AND MAINTENANCE

F-1.

General

This section of the manual contains routine
maintenance information regarding the
AMPTEC RESEARCH 601ES Safety DMM.
Calibration should be performed on a regular
basis to ensure continued instrument accuracy
or following a main PCB electronic component
repair/replacement. The recommended
calibration interval is 1 year.

in order to eliminate lead resistance

and contact resistance errors. The Option “300”
is a gold plated 4 Wire Kelvin Clip Test Lead
set that is available with the AMPTEC 601ES
DMM.

AMPTEC 601ES DMM Calibration note:

verify @ 60Hz

For the resistance function, the AMPTEC
601ES Safety DMM must be calibrated using
four wire Kelvin connections to the resistance
standard

F-2.

Required Test Equipment

Following standard resistors are required to
calibrate the 601ES Explosive Safety Meter
Igniter Tester .

1 ohm ± 0.01% Accuracy

10 ohms ± 0.005% Accuracy
1000 ohms ± 0.005% Accuracy

10 Kohms ± 0.01% Accuracy

600 V ± 0.05% or better accuracy
100 V ± 0.05% or better accuracy

10 V ± 0.05% or better accuracy

1.0 V ± 0.05% or better accuracy

600 V ± 0.05% or better accuracy
100 V ± 0.05% or better accuracy

10 V ± 0.05% or better accuracy

1.0 V ± 0.05% or better accuracy

1.0 ampere ± 0.05% or better accuracy

Precision Resistance Standards:

DC Voltage Standard:

AC Voltage Standard:

AC/DC Current Standard:

F-3. Calibration Procedure

F-3-1. Zero Offset Adjustment

1.

2.

A.

DMM

Function and

Following a full 15 minute warm-up

period

Only

Out of Specification

Section B

No adjustments are needed

-

- Verify each function and range with the

appropriate calibration standards to determine if
the AMPTEC 601ES DMM meets its published
accuracy specifications as published in Section B
( p.7).

- If the 601ES meets the specifi-

cations as compared with traceable standards then
it is ready to certify with a calibration sticker etc.
and "

".

its the PCB directly under

the "

" functions

(

) need to be addressed with any adjust-

ment procedure

The 601ES

DMM requires fully charged batteries and a full 15
minute warm-ip period prior to any adjustments.
The calibration adjustments are accessed by
removing the 10 calibration access screws around
the perimeter of the 601ES

front panel, then

lifting off the top plate by the handles. The
locations of the adjustments are shown on drawing
number 601ES Explosive Safety Meter-600 at the
back of this manual.

Any zero offset greater than the specification (

0.05% of range which is 00.10 on the 20 Kohm
range) then adjust as noted

(see figure H-14 in the back of the manual for

the DVM PCB location -

With the AMPTEC 601ES DMM in 4 Wire mode,

select the 20 Kohm range. Short the four Wire
Kelvin Leads together at the ends.

next. If the zero reading

is >00.10 then adjust potentiometer RV405 to be as
close to 00.00 as possible ( its located on the DVM
PCB)

Verification of Each

Ranges

Accuracy

Use caution when lifting the

601ES top panel up by the handles

, as there is

umbilical wiring that connects the top panel
electronics to battery pack wiring and chassis/case
ground wiring mounted inside the bottom of the
601ES case. Set the top panel on end resting on the
unit’s bezel.

19

Содержание 601ES

Страница 1: ...RESEARCH AMPTEC CALIBRATED MODEL SER NO DATE DUE Tamper ProofSeal Sticker Calibrationvoidifremoved 601ES 601ES 503 05APR2005 05APR2006 I press hold 2 0 KOHM 20 OHM 200 OHM 2 0 OHM 20 KOHM FUNCTION RA...

Страница 2: ...cifications C 1 Available Accessories and Options D 1 General D 2 Front Panel Features and Operation D 3 Complete 601ES Explosive Safety DMM Front Panel Diagram D 4 4 Wire Resistance Measurement 601ES...

Страница 3: ...Procedure F 3 1 Zero Offset Adjustment F 3 2 Full Scale Adjustment F 3 3 2 0 Ohm Range Adjustments F 3 4 Verification of Each Ranges Accuracy F 3 5 AC DC Voltage Adjustments F 3 6 AC DC Current Adjust...

Страница 4: ...osive Safety Igniter Circuit Test industry must be held down for 2 seconds The AMPTEC 601ES DMM Safety Meter has a circuit that indicates when it is time To replace the batteries They can be replaced...

Страница 5: ...ince Nickel Metal Hydrides are re chargeable they offer a battery life cycle a hundred times greater than any alkaline Be sure to charge the Nickel Metal Hydride batteries when taken out of their orig...

Страница 6: ...around the battery holder and secure Cut excess plastic Next wrap the black and red wire around the battery holder and place the holder into the compartment Lastly replace lid and screw in screws to...

Страница 7: ...ange Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range DC or AC RMS range Accuracy 0 50 of range 2 0 Amperes DC or AC RMS range Accuracy 0 50 of range 2 0 Ohm Range thru 20 KOhm ranges...

Страница 8: ...e point of the probe tip connection eliminating 98 of the cable resistance Gold plated tips and banana jack ends provide an optimal low thermal emf connection gold plated single points gold plated gol...

Страница 9: ...banana plugs at one end The other end is terminated with an alligator clip red and black These alligator clips will measure 2 wire method properly with either upper or lower jaw connected to the condu...

Страница 10: ...once the access screws have been removed for calibration These handles are for carrying the 601ES DMM around when the lid is removed They are strictly for calibration access he main Power switch to tu...

Страница 11: ...M is factory set using internal PCB jumpers to automatically turn off after 30 minutes of inactivity no range or function change to conserve battery power 5 Automatic Turn Off Protection Fuse Stray Cu...

Страница 12: ...g the zero knob past the stops will severely damage the precision of the instrument and violate the instruments warranty 2 20 and 200 ohm ranges 10 Case Chassis Ground Jack not tied to earth ground A...

Страница 13: ...s go in continuous decades from 2 0 Ohms to 20 Kilo Ohms The selection of the 2 0 Ohm lowest resistance range is accomplished by rotating the range knob Align indicator mark on knob with the desired r...

Страница 14: ...ck to the AMPTEC 601ES DMM through the jack Depending upon the magnitude of the resistance offset some of the offset may be corrected zero ed out for in the AMPTEC 601ES DMM by adjusting the zero knob...

Страница 15: ...g Circuitry fatigued adapter wire plug wire splice loose the fewer the connections the fewer the problems The LCD display will indicate when it is time to replace the batteries LO BAT LO BAT The The A...

Страница 16: ...atsoever If you turn on display does not come on Explosive Safety DMM m AMPTEC Explosive Safety DMM and the If the AMPTEC 601ES Explosive Safety DMM exhibits problems that cannot be eliminated by revi...

Страница 17: ...only when the integrator is de integrating At the end of a conversion the data remaining in the results counter is latched decoded and multiplexed to the liquid crystal display The analog section blo...

Страница 18: ...n also turn the unit off the same way after a certain time period has elapsed This is to save battery life See FIG H 1 The micro can turn on TR201 TR209 using IC201 and I 202 These fets are used to se...

Страница 19: ...y Out of Specification Section B No adjustments are needed Verify each function and range with the appropriate calibration standards to determine if the AMPTEC 601ES DMM meets its published accuracy s...

Страница 20: ...PCB then later verify 3 4 5 6 7 8 A B C D F 3 6 AC DC Current Adjustments 1 the battery compartment When adjusting RV405 which is indicated as a negative sign Select the 20 Kohm range Connect the Kel...

Страница 21: ...press to measure button and Adjust RV502 and repeat this step 5 dual button press operation until the AMPTEC 601ES reads 0050 A 5 milliamperes Return both the and the s to their standoff PCB mounts Re...

Отзывы: