AST 560 Semi-automatic Cryostat Microtome 

201911119 

---19--- 

and the base holder locking lever and the storage area on the crust. Use the dry cloth to clean 
other appearance. 

 

Cleaning the clamp

 

As showed in fig 23.

Specimen clamp position (1) which is  often touched  when  operation,  and 

position  (2)  which  always  contacts  with  the  specimens.  These  two  positions,  especially 
position (2), are very easy  to  be  polluted, so  need to  be  cleaned frequently  to  ensure normal 
operation. 

       

 

Fig 23                                                                          Fig 24 

 

Cleaning the blade holder

 

T

ake  down  all  the  parts  of  blade  holder  as  it  is  showed  in  the  picture  and  then  clean  all  the 

parts  separately,  especially  the  following  easily  polluted  parts:  the  sliding  guide,  fixed  pole, 
blade  clamp  and  the  joint  of  the  parts.  And  remember  to  clean  the  blade  clamp  every  time 
before installing the blade to ensure good section. 

To obtain a high quality section, it is important to keep cleaning the instrument. So the user 

must clean the instrument periodically or irregularly according to the total sectioning quantity to 
obtain the best section. 

  Only  authorized  and  qualified  service  personnel  may  access  the  internal  components  of  the 

instrument for cleaning and maintenance .                                             

  Before cleaning and maintenance turn off the instrument , pull out the plug and take down the 

blade holder and then clean all the parts of instrument separately . The blade must be taken 
down before cleaning the blade holder.                                                     

  Lock the handwheel before cleaning .                                                                           

  Open the glass after turning off the instrument to keep the chamber dry !                                                                           

  Do not use any corrosive solvent for cleaning !                                                                     

  Ensure no liquid enters the interior of the instrument during cleaning !                                 

  Keep cleaning the cryo chamber frequently !                                                                         

  Do not turn on the instrument before it is completely dry .                                                   

                             

                           

5.2 

Cleaning the waste tissue

 

Section extraction is equipped which can conveniently remove the waste tissue produced by 

Содержание AST 560

Страница 1: ...AST 560 Semi automatic Cryostat Microtome Operation Manual www amos scientific com AMOS SCIENTIFIC PTY LTD ABN37 159 778 140...

Страница 2: ......

Страница 3: ...Presented in this Operation manual are the structure functions and using notice Reading carefully prior to operating the instrument...

Страница 4: ...ly it is necessary to read the operation manual carefully Our companies are not only responsible for the repair but also train distributors ability of repairing To repair the product smoothly in time...

Страница 5: ...INSTALLATION 6 3 4 ELECTRICAL CONNECTION 7 4 OPERATION 8 4 1 OPERATION PANEL FUNCTION CONTROL 8 4 3 AUXILIARY FUNCTION 10 4 4 CRYO CHAMBER TEMPERATURE SETTING 11 4 5 SPECIMEN CLAMP TEMPERATURE SETTIN...

Страница 6: ...the factory please check this setting complies with your local power requirement before connecting to the main power Please use the power cord provided If need to replace it please use the power cord...

Страница 7: ...cord before changing the fuse Please use the fuse according to the requirement and specification in operation manual 1 3 Safety Device Handwheel locking mechanism As showed in Fig1 it is the lock lev...

Страница 8: ...sing etc which is for the pathological analysis of tissue samples 2 2 Overview Instrument Components The main structure of cryostat microtome is shown in Figure 3 as below 1 Touch Panel 7 Heating glas...

Страница 9: ...on is enabled or disabled Defrosting at fixed time or manually Sterilization at fixed time or manually Sleep at fixed time or manually Large freezing shelf could load 17 samples simultaneously Quickly...

Страница 10: ...wo fixed support feet and make the front wheels disabled Start the instrument after let it rest for at least 2 hours Ensure no other goods around the instrument At least 300mm distance must be kept fo...

Страница 11: ...in order Upholder 2 Corner Support 1 3 Corner Support 2 6 Corner Support 3 7 Corner Support 4 11 Handwheel 8 Blade Holder 9 Cryostat Microtome 12 Glass Support 5 Then take out other accessories in th...

Страница 12: ...check that this setting complies with the local power requirements Connect the instrument to a ground power socket to avoid accident Please use the power cord provided If need to replace it please us...

Страница 13: ...he icon pop up sub interface select the auxiliary function 3 Lamp Click the icon turn on off the lamp 4 Time display and set Display current date and time Click the icon set the date and time 5 Cryo c...

Страница 14: ...number sum Section number sum when exchange between slicing and trimming the value will return to zero automatically 14 Progress display Display the sample clamp progress 15 Section thickness set Acc...

Страница 15: ...ate to continue setting 3 Right arrow Use cooperatively with left arrow 1 select one date and set its working time for each function 4 Automatic defrosting time Set the selected date and set the autom...

Страница 16: ...an 8 click this icon to activate this function It will work for 15 minutes and stop 5Pump waste sweeps Prepare the related tools click on this icon the device automatically pump cutting waste sweeps 6...

Страница 17: ...Value adjustment Double arrows can be adjusted once 5 single arrow can be adjusted 1 3Quick value selection Select temperature value quickly 4Back After setting the parameters click it back to the mai...

Страница 18: ...n click SECT TRIM icon once it turns to retraction mode while TRIM turns to green When the current mode is retraction RETRA is green click RETRA icon once it turns to white while SECT turns to green w...

Страница 19: ...ction status the icon is highlighted or it is dark 6 Heating glass opened When the heating glass is open the icon is highlighted or it is dark 7 Alarm When the device has alarm information the icon is...

Страница 20: ...tion of blade holder and base As showed in fig 17 slide the blade holder along the track in cryo chamber turn the fixed lever 2 to horizon position to lock the blade holder The blade is close to speci...

Страница 21: ...e and put it into box when not in use Fig 17 Adjust the angle As showed on fig 18 the rotating angle is from 0 to 10 degree the user can adjust the angle in this range according to requirement Fig18 F...

Страница 22: ...er 2 Always clamp the specimen before installing the blade to avoid injury Fig 20 Blade inserting As showed in fig 21 loosen the fixed lever 5 and push the blade 6 into blade holder 4 then rotate the...

Страница 23: ...ntion to the following aspects 1 Proper cryo chamber temperature selection 2 Correct section operation 3 careful anti roll plate adjustment 4 Sharp blade and appropriate sectioning angle 5 The blade i...

Страница 24: ...ction To obtain a high quality section it is important to keep cleaning the instrument So the user must clean the instrument periodically or irregularly according to the total sectioning quantity to o...

Страница 25: ...e white tie tightly along the edge to fix firmly Fig 26 Fig 25 Fig 26 Pull off the silica plug of sucking hole as the Fig 27 and properly placed Fig 27 Plug the adaptor with filter bag into the suctio...

Страница 26: ...and then removed the scrap bag to clean the waste scraps or change a new scrap bags to continue work 5 3 Maintenance Fuse replacement Put the fuse 1 in the installation part 2 and then insert the ins...

Страница 27: ...than the blade Adjust the anti roll plate Section soften The temperature of blade or anti roll plate is too high Extend the cooling time of blade or anti roll plate The section sticks to the blade or...

Страница 28: ......

Страница 29: ......

Страница 30: ...Published by AMOS SCIENTIFIC PTY LTD Address 1 Prestige Dr Clayton South VIC 3169 Melbourne Australia Tel 61 4 10190188 Fax 61 3 98081503 Website www amos scientific com Email info amos scientific com...

Отзывы: