AST 560 Semi-automatic Cryostat Microtome
201911119
---15---
correct position and fix it by fixed lever 3.
⊙
Fixed lever 4: Rotating movable jaw can be moved in the direction of arrow A on longitudinal
movable jaw to adjust the holder angle. When it reaches to correct angle, fix it by lever 4
⊙
Anti-roll plate height adjusting knob: Turn the adjusting knob to adjust the relative distance
between the anti-roll plate and the blade, to adjust the anti-roll affection.
The blade holder has been already assembled when leaves factory, it only needs to be installed in
the equipment.
Fig 16
Installation of blade holder and base
As showed in fig 17, slide the blade holder along the track in cryo chamber, turn the fixed
lever 2 to horizon position to lock the blade holder. The blade is close to specimen clamp, please
do as the steps:
⊙
If there is bladed installed, please take care to remove the blade and place it properly.
⊙
Turn the fixed lever 2 (2) slowly until the locking block (3) at the bottom of the blade holder
moves to the farthest position away from the blade holder.
⊙
Aligning the locking block (3) at the bottom with the circular hole (4) of the rail(1) ; Meanwhile ,
align the V-shaped groove (5) at the bottom of the blade holder with the V-shaped groove (6) on
rail (1) , and then place the blade holder on the rail (1) gently and push it forward a reasonable
3
4
5
6
7
8
10
9
11
2
1
10
A
B