AST 560 Semi-automatic Cryostat Microtome
201911119
---1---
1
、
Safety Notes
1.1Overview
The Operator Manual contains important safety instructions and information. In order to
operate the instrument safely, the operator must read carefully prior to starting up the instrument.
It is very important to ensure safety and avoid instrument damage. Please keep the manual close
to the instrument to read easily at any time.
▲
Note:
The safety marks and protective device on both instrument and accessories may
neither be removed nor modified , so as not to injure human body or instrument .
1.2 Safety Warning
Every principle person should read carefully and carry out strictly according to the following
safety notes relating to transport, installation, calibration, operation, maintenance, cleaning and
all aspects of the instrument.
1.2.1 Warning during transport and installation
●
The instrument may only be transported or moved in an upright position and the tilt
angle should not exceed 45
°
.
●
Prior to removing the microtome , remove the blade holder .
●
The input voltage has been preset at the factory , please check this setting complies with
your local power requirement before connecting to the main power
●
Please use the power cord provided . If need to replace it, please use the power cord
with earth wire.
●
Do not operate this instrument in the room with the danger of explosion .
●
Don't remove or modify the protective device on both instrument and accessories , in
order to avoid injury to instrument or human body !
1.2.2 Warning of Operation
●
Take care when handling microtome knives and disposable blades . The cutting edge is
extremely sharp and might cause serious injury.
●
Remove the knife/blade before detaching the blade holder from the instrument , and put
the blade back into the blade box when not in use
●
Never place the blade with the cutting edge facing upwards and never take out the blade
with your hands.
●
Please clamp the specimen block before clamping the blade .
●
Prior to replacing the specimen and blade , please lock the handwheel and cover the
blade guard on the cutting edge .
●
Rotate the handwheel clockwise, otherwise it would affect section .
●
Don't rotate the handwheel clockwise and counterclockwise frequently at top and
bottom , it might lead to deviation of section thickness .
●
Ensure no liquid enters into interior of instrument during working .