AST 560 Semi-automatic Cryostat Microtome
201911119
---11---
Click the auxiliary function setting icon on the left side of the main screen to pop up the
auxiliary function sub-interface as below fig 11
.
Fig 11
1Real-time defrosting Click this key to enter defrosting status immediately
2Real-time
disinfection
Click this key to enter disinfection status immediately. Please note it
disinfects only when the heating window closed
3Real-time sleep
Click this key to enter sleep status immediately
4Specimen clamp
peltier switch
When the specimen clamp temperature is lower than -8
℃
, click this icon to
activate this function. It will work for 15 minutes and stop.
5Pump waste sweeps
Prepare the related tools, click on this icon, the device automatically pump
cutting waste sweeps
6 Feed to memory
position
When the sample feeding, memory the current position and use
cooperatively with item 7.
7Automatic feeding Click this icon, it feeds to memory position automatically.
8Alarm information
When the device alarms, click it to check the alarm information.
9 Back
After setting the parameters, click it back to the main interface.
Notice:
Be sure to take out the samples and waste bags etc before entering into defrosting status.
The item 6 and 7 is only suitable for samples with similar thickness.
4.4
Cryo chamber temperature setting
1
2
3
4
6
5
7
8
9
10