AEM460 Semi-automatic
Microtome
AMOSAEM460OP20180306
---16---
●
Take care when operate the knife holder and the blade. The cutting edge is
extremely sharp and can cause serious injury.
4.5 Trimming the specimen
⊙
Retraction setting
Active retraction function and set retraction thickness via the
button.
Press to light up the LED
,
while the retraction
function be activate, and deactivate by press again to light
down.
Long press the about three sec. Till LED light up
and flicker, the others button invalid, while the user can select the
retraction value via , the value will display on .To exit
the setting, push the others button TRIM or SECT.
⊙
Trimming setting
Press button while LED light up
,
the trimming
function be activate. Press to select the trim thickness
value displayed on .
⊙
Adjust the specimen position.
Press these buttons to adjust the specimen back and forth to
ensure the specimen is parallel with the cutting edge and wait
trimming.
Press to adjust when the specimen is far away to cutting
edge. It is fast.
Press to adjust when the specimen is near to cutting edge
⊙
At last
,
check and ensure the three levers on the knife holder,
lever on the base holder and the lever on the specimen clamping
system all be locked. And then unlock the right handwheel to
trim.
●
Before change the specimen and the blade, the specimen
clamping system must be centralized locked in upper.