4.3
Clamping the Specimen
As the drawing shows , turn the
handwheel (2) to drive the specimen clamp(1)
moving in the vertical direction.
●
To clockwise turn the handwheel when
sectioning to obtain a best section.
As drawing shows, the specimen
clamping system is made up of adjustor(3)
、
connect board (4) and clamp(5
、
6).
There are two kinds of clamp: Specimen
clamp (6) and Quickly release cassette clamp
(5).
Specimen clamp is fit for
:
Dimension of specimen:40
*
50*30mm or
the standard cassette.
Quick release cassette clamp is fit for:
Standard cassette.
Use the screws to fix each component as
Fig shows. To change the specimen clamp, the
user can loosen the screw connected the
specimen and connector.
The specimen clamp adjustor
Loosen the lever 1 (7)
,
and adjust the vertical
knob (8) to calibration the vertical angle of the
specimen; and the horizontal adjusting knob(9)
can be used to adjust the horizontal angle of
the specimen.
After the above, use rotating
lever 1 (7) to
tighten it.
Specimen adjusting system:
horizontal orientation : ±8°
vertical orientation
:
±8°
1
2
6
4
5
3
8