AEM460 Semi-automatic

 

Microtome

 

AMOSAEM460OP20180306

 

---5--- 

2.2 

Performance Index

 

AEM460  is  a  motorized  rotary  microtome.  The  stepper  motor  operated  the  specimen 

movement forward and back, coarse feed . It is better in precision sectioning, easy operation. 

  The  crust  is  made  in  ABS  plastic  with  fire  prevention  function,  and 

streamlined

 

designed in artistic appearance. 

  Spacious 

ingenious assembly

 section waste tray. 

  Retraction function make it is better to sectioning. The retraction value can be set. 

  Quick  release  cassette  clamp  and  specimen  clamp  fit  to  internal  Embedding 

cassette. 

  Locking  may  be  activated  at  any  position  by  handwheel.  It  ensure  cassette 

changing safety when the cassette located in the uppermost locked. 

  The electric coarse feed operates two speeds which is easy to record the feed. 

  Separate control panel is more easy to operate. 

  Automatic  memorized  coarse  feed  and  reload  fast  can  improve  efficiency  during 

sectioning. 

 
 

2.3 

Technical Data

 

  Surroundings requirements:   

Working temperature

+10℃—40℃

 

Working humidity: <80%

defrosting   

Working pressure:

86

106

kPa

 

  Power supply

100-240 V AC±10 % 

  Frequency

50/60 Hz 

  Power

        <60 VA 

  Fuse

    2A 

  Safe classify

Classify

- type B   

  Section thickness

  0 to 600µm 

                                        0 to 2µm

0.5µm increments 

                                        2 to 10µm

1µm increment   

                                        10 to 20µm

2µm increments 

                                        20 to 100µm

5µm increments   

                                        100 to 600µm

50µm increments 

  Trimming thickness:    0 to 600µm 

                                        0 to 2µm

0.5µm increments   

                                        2 to 10µm

1µm increment 

                                        10 to 20µm

2µm increments 

                                        20 to 100µm

5µm increments   

                                        100 to 600µm

50µm increments 

  Retraction thickness: 5-100µm. 5µm increments (can deactivate)   

 

Specimen horizontal feed:28mm 

Содержание AEM 460

Страница 1: ...AEM 460 Version AMOSAEM460OP20180306 Semi automatic Microtome Operation Manual www amos scientific com AMOS SCIENTIFIC PTY LTD ABN37 159 778 140...

Страница 2: ......

Страница 3: ...and using notes Be sure to read these instructions before using the instrument Quick release cassette Clamp Knife holder Display Panel Separate control panel Handwheel Locking Mechanism Section Waste...

Страница 4: ...ent is snuff long haul and safely used Our company is responsible for service after the microtome have been used our agents have ability to service the client all of world can connect with our agent i...

Страница 5: ...6 3 1 Installation Site Requirement 6 3 2 Standard Delivery 6 3 3 Installation 7 3 4 Electrical Connection 8 4 Operation 9 4 1 Control Panel Function Control 9 4 2 Knife Holder inserting 12 4 3 Clamp...

Страница 6: ...strument 1 2Safety Warning The following safety notes relating to transport installation calibration operation maintenance cleaning and all aspects of the instrument Every principal person must be rea...

Страница 7: ...e knife guard when slicing Don t Clockwise and Counterclockwise rotate the handwheel frequently at top and bottom Slice thickness many deviations Ensure that liquids don t enter the interior of the in...

Страница 8: ...2 is not activated When the lock is activated the LOCK indicator will light on display panel while the machine DON T running DON T lock the handwheel when the handwheel is rotating otherwise it will...

Страница 9: ...assette Clamp Display Panel Handwheel locking Mechanism lock at fixed position Handwheel Locking mechanism lock at any position Knife Guard at knife holder Knife holder Clamping lever for knife holder...

Страница 10: ...ted in the uppermost locked The electric coarse feed operates two speeds which is easy to record the feed Separate control panel is more easy to operate Automatic memorized coarse feed and reload fast...

Страница 11: ...around the instrument to avoid the shelter affecting the operator operate the instrument The working temperature and humidity must be accord with the data in the manual Ensure there is enough space f...

Страница 12: ...4 specimen clamp 7 front protect board 6 back protect board 3 and then the instrument 2 in turn Final to take out other accessories in the inner package to install the instrument 3 3 2 Knife holder Ta...

Страница 13: ...ocal power requirements of your laboratory Connect the instrument to a grounded power socket to avoid accident It is better to use the power cable provided or use an extension cord with a ground wire...

Страница 14: ...isplays Current set value of section Thickness trimming section thickness setting The second line displays section trimming counter and section thickness sum The trimming status stand by when instrume...

Страница 15: ...lit up it shows the status at trimming and trimming Thickness setting SECT If its Led lit up it shows the status at sectioning and section Thickness setting RETRA If its Led lit up it shows the statu...

Страница 16: ...m s Specimen feed is 28mm If exceed the value the buzzer will alarm with voice and the specimen movement stopped Button for trimming Light up when trimming mode is active Trimming and trimming thickne...

Страница 17: ...and then adjust the adjustable handle of the lever 2 7 to horizontal position Rotate the rotating X axis slider 3 to slide it into the Y axis slider 2 and then insert the lever 2 8 and rotate it until...

Страница 18: ...e knife before detaching the knife holder from the instrument Always put the knives back into the knife case when not in use Tighten the knife holder and the base holder by the four levers and adjust...

Страница 19: ...n clamp is fit for Dimension of specimen 40 50 30mm or the standard cassette Quick release cassette clamp is fit for Standard cassette Use the screws to fix each component as Fig shows To change the s...

Страница 20: ...the specimen 6 into and then free loosen lever 5 the specimen will be clamped automatically To take down the specimen just operate as above Easy to change the specimen by one hand operating the quick...

Страница 21: ...the setting push the others button TRIM or SECT Trimming setting Press button while LED light up the trimming function be activate Press to select the trim thickness value displayed on Adjust the spec...

Страница 22: ...se standard specimen holder and standard cassette without any exceed paraffin 2 Don t move the knife holder during the whole operation 4 6 Sectioning Sectioning setting Press button while LED light up...

Страница 23: ...g and position 2 which is often contacted with the specimens especially position 2 are very easy to be polluted so need to be cleaned frequently to ensure ordinary operation Quick release cassette cla...

Страница 24: ...cess the internal components of the instrument for clearance and maintenance before clearance and maintenance turn the instrument off pull out the plug and take down the knife holder and then clean al...

Страница 25: ...the specified type fuse Maintenance for the knife holder Lever the fixed installations for the knife holder and other parts of the instrument which are frequently used and easily worn need to be maint...

Страница 26: ...too small The clamping setting is not steady The paraffin is too hard The angle between knife and the embedded cassette is too large The set sectioning thickness is too Thick Blunt cutting edge Check...

Страница 27: ...st situations Trim the embedded specimen to make the top line parallel with the bottom line and the width of the embedded specimen must be even Adjust the specimen clamping to ensure the top line and...

Страница 28: ...araffin too long or the paraffin is too hot The specimen is too hard to the paraffin The blade is uneven The blade is polluted The specimen is scratched by hard particles There are calcareous or silic...

Страница 29: ...olved please contact with the manufacturer 7 Instrument Diagram The surface of the section is in wave The section flies away and sticks to the microtome or other objects near the microtome The gradien...

Страница 30: ...1 unit 2 Blade Holder 1 set 3 Specimen Clamp 1 set 4 Waste Tray 1 pc 5 Separate control panel 1 set With connection cable 6 M3 Allen Wrench 1 pc 7 M4 Allen Wrench 1 pc 8 Disposable blade 1 box 9 Dust...

Страница 31: ...Published by AMOS SCIENTIFIC PTY LTD 7 Gabrielle Court Bayswater North VIC 3153 AUSTRALIA Tel 61 4 10190188 Fax 61 3 98081503 ABN 37 159 778 140 www amos scientific com Email info amos scientific com...

Отзывы: