User Manual for AML Oceanographic Minos SVP
3
Shipping & Receiving
Receiving an Instrument
When an instrument is received at a new location it is prudent to perform the following steps to
ensure the instrument is capable of performing when required.
•
Check the shipping container for signs of damage
•
The shipping package should include all of the following items
o
Minos SVP instrument
o
Data/Power cable
o
Red shorting plug
o
Black dummy plug
o
Manual
o
Battery charger
•
Check for damage
o
Check the cable for slices or gouges
o
Check the connectors for corrosion in sockets and salt deposits
o
Check the pressure case for dents and scrapes
o
Check the sensors for cracks or bends
•
Check tightness of connection of any swappable sensors (see the Xchange™ manuals for
more details on required tightness)
•
Connect the instrument to a computer with the data cable and perform a scan or monitor if
using SeaCast. Check the battery voltage on the output (normally the last column of data
unless calculated parameters have been enabled). It should be between 7.1 and 8.4 volts.
Returning an Instrument to the Factory
•
If shipping for repair or recalibration obtain an RMA number from the service centre.
Please refer to the section Contacting AML Oceanographic for contact details.
•
Perform all the steps listed above for Receiving
•
Pack the instrument to prevent damage during shipping