
User Manual for AML Oceanographic’s Micro SV & P Sensor
9
4
PREPARATION FOR USE
4.1
INSPECTING THE INSTRUMENT
At AML we do our best to package our instrumentation to avoid damage during shipping,
as should the user. However accidents do happen, so an inspection of the Micro SVP
Sensor before each use will assist in spotting problems that could lead to inaccurate data or
possible failure. Examine the outside of the shipping case for evidence of heavy impacts
during transport. If signs of damage are visible continue with the inspection as follows and
notify the carrier and the factory of any damage found. Check that the communications and
power connector is not loose and that there is no dirt or grit in the connector(s). Examine
the cable for cuts or wear and check the connector ends for visible damage. Also check for
a cracked or chipped sound velocity sensor, as this will change the calibration of the sensor.
4.2
CONNECTION TO A COMPUTER
The user communicates with the Micro SVP Sensor via any IBM or compatible computer or
data collection device. An AML supplied data cable will link the two. Refer to the wiring
diagram in
Appendix C
. At one end of the cable is a DB25 or DB9 female connector that
plugs into the computer's serial port and at the other end is an 8 pin plug that is inserted into
the communications port of the Micro SVP Sensor. When this connection has been made
and 12 volts applied, the instrument is powered up and ready to communicate with the
computer.
WARNING:
A plug or cable must be installed in the connector at all times when the instrument is
immersed in water. Failure to do so will cause corrosion to the connectors and may
cause water damage in the electronics housing
.
4.3
POWERING THE MICRO SVP SENSOR
4.3.1 External Power
The Micro SVP Sensor is normally powered externally via the communications cable. Refer
to the wiring diagram in
Appendix C
. If an AML data cable is used attach the red wire to
the positive side of the power supply and the black wire to ground. The power should be a
DC voltage between 8 and 16 volts, measured at the instrument. The instrument has reverse
voltage protection but the instrument will not function if the power leads are connected with
reverse polarity.
Figure 4.3.1
Micro instrument connected to an external power source