SVP Micro Sensor V2.01 SN:1234-SVP
Copyright(C) 2003, AML Oceanographic
32MB of Memory
Warning: Changing the instruments serial number will adversely effect the
operation of Smart Talk or ISS software.
Command:
SET DETECT ab [SE D ab]
Where
a = a Hex number between 0-F
b = a Hex number between 0-F
This command sets the detection mode of the instrument on power up.
The DETECT command can be used to set the Micro SVP to start up in the
Auto baud
mode
or to set the instrument to default to a specific baud rate at power up.
The ‘a’ value represents how many times the Micro SVP will try to determine the baud
rate (when the enter key is pressed) before it defaults to the baud rate set by the value of
‘b’. If a = 0 the Micro SVP will not auto baud and will default to the baud rate specified
by the value of ‘b” at each power up.
Specific values of ‘b’ and the corresponding baud rates are outlined in the table below.
b value
Baud rate
1
1200
2
1200
3
2400
4
4800
5
9600
6
19200
7
38400
Example 1:
>set detect a7
In the above example, the Micro SVP will try 10 times to auto baud. If the
instrument is not able to establish a baud rate, it will default to 38400 baud.
Example 2:
>set detect 05
In the above example, the Micro SVP will
not
try to auto baud, but will simply
default to 9600 baud on power up.
Command:
SET DEFAULT [SE DEF]
This sets the instrument to its original factory default settings.