User Manual for AML Oceanographic’s Micro SV & P Sensor
24
Example:
>talk 203
(enter talk mode with the sound velocity board)
Entering talk mode
Sound Velocity Micro Sensor V1.40 SN:0002-SV
>set a=7.361894E-04
(set the a coefficient)
>dis c
(display the sound velocity sensor coefficients)
Sound Velocity
A=7.361894E-04 B=-7.438164E-056 C= 1.07834E-06 D=-5.791932E-07
Press ctrl-c to exit the talk mode.
Pressure
Command:
SET Pn=c
Where n = coefficient letter a,b,c,d,e,f,g,h,i,j,k,l
c = calibration coefficient
Example:
>talk 200
(enter talk mode with the Pressure and Temperature board)
Entering talk mode
Pressure and Temperature Micro Sensor V1.30 SN:0001-PT
>set pa=1.00
(set the a coefficient)
>dis c
(display the calibration coefficients)
Pressure
A= 1.00 B= 1.00 C= 0.00 D= 0.00
E= 0.00 F= 0.00 G= 0.00 H= 0.00
I= 0.00 J= 0.00 K= 0.00 L= 0.00
Temperature
A=0.00 B=0.00 C=0.00 D=0.00
E=0.00 F=0.00 G=0.00
Press ctrl-c to exit the talk mode
Analog Board
Refer to
Appendices A
and
B
for instrument channel configurations.
Analog Channel 1
Command:
SET 1n=c
Where n = coefficient letter a,b,c,d
c = calibration coefficient