User Manual for AML Oceanographic’s Micro SV & P Sensor
21
Micro
SVP
power up timing diagram
Command:
SET STARTUP HEADER [SE ST H]
Display the header at power on.
Command:
SET STARTUP NOHEADER [SE ST NOH]
Do not display the header at power on.
Command:
SET STARTUP MONITOR [ SE S M ]
This command sets the Micro SVP to automatically output scans at power on.
If the auto baud feature has not been disabled, the Micro SVP will wait for a
carriage
return
before it outputs data. (See SET DETECT command).
The following example set the Micro SVP to output data at power up with a baud rate of
9600.
>set detect 05
(disable the auto baud and default to 9600 baud at power up)
>set startup noheader
(do not display a header at power up)
>set startup monitor
At power up the Micro SVP will continuously output Real data at 9600 baud.
Note: To return to the prompt, hold down the RETURN key while powering up the Micro
SVP. Use the SET STARTUP PROMPT command to return the Micro SVP to the
prompt mode.
Command:
SET STARTUP SCAN [SE S S]
This command sets the Micro SVP to automatically output one scan of data at power on.
If the auto baud feature has not been disabled, the Micro SVP will wait for a
carriage return
before it outputs data. (See SET DETECT command).