User Manual for AML Oceanographic’s Micro SV & P Sensor
10
4.3.2 Micro SVP Logging Option
If the Micro SVP has been purchased with the logging option it will be equipped with
internal memory and an external battery pack. Refer to the wiring diagram in
Appendix C
.
If the battery pack is used, the instrument is connected to either connector on the battery
pack with an interconnect cable. The remaining connector on the battery pack is then
connected to the data/power cable, which is connected to the computer. For logging data
when not connected to the computer the red shorting plug must be inserted into the battery
pack to turn the power on to the instrument. Power for the Micro SVP sensor is turned off
if the there is no shorting plug or data/power cable plugged in to the battery pack.
Figure 4.3.2
Micro instrument with battery pack and external sensor
connected to an external power supply
4.3.2.1 Micro Battery Pack
Figure 4.3.2.1
The Battery pack
The replacement batteries should be three, good quality, C size, lithium, 3.6 volt batteries.
Saft, LSH 14, C cell, 3.6 volt, lithium batteries are recommended. For more details on
battery recommendations refer to
Section 7.2
.
Insert all three batteries into the brass housing positive end first. Replace the battery cap on
the brass housing.
Check the O-ring seals. The O-rings should be clean, have no nicks, and should be lightly
greased. Slide the battery assembly back into the pressure case and screw down the
retaining ring.