User Manual for AML Oceanographic’s Micro SV & P Sensor
25
Analog Channel 2
Command:
SET 2n=c
Where n = coefficient letter a,b,c,d
c = calibration coefficient
Example:
>talk 202
(enter the talk mode with the analog board)
Entering talk mode
Analog Micro Sensor V1.30 SN:0003-A
>set 1a=1
(set the a coefficient of channel 1)
>set 2a=2
(set the a coefficient of channel 2)
>dis c
(display the calibration coefficients)
Channel 1
A= 1.00 B= 0.00 C= 0.00 D= 0.00
Channel 2
A= 2.00 B= 0.00 C= 0.00 D= 0.00
Press ‘crtl-c’ to exit the talk mode.