Process Pressure Checks (Only if above 2 psig)
To check that you entered the process pressure correctly, select
Process Pressure
from the
configuration menu
. Also ensure that the calibration process pressure equals normal operat
-
ing process pressure. Calibration should only be performed under these conditions for highest
accuracy.
Cell Checks
If the cell fails when you first begin
to use the analyzer:
It is likely that there is a leak in the sensor plumbing
or an improper calibration gas setup, and there is not
a problem with the cell itself (see the “Leak Check”
section later in this chapter for help on how to check
for leaks; see the “Calibration/Aspirator Air Checks”
section earlier in this chapter for help on checking
your calibration gas setup).
If the analyzer has been operating
for some time and you feel the oxy-
gen reading is inaccurate:
First check by running a known calibration gas to veri-
fy the analyzer’s response. If the analyzer responds to
the calibration gas correctly, it indicates either leaking
or plugged plumbing.
If your analyzer doesn’t respond
properly to the calibration gas:
This may indicate a problem with the cell. Before re-
placing the cell, check for leaks or plugged plumbing.
Содержание Thermox WDG-V UOP
Страница 34: ...2 6 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...
Страница 65: ...AMEVision Display Installation Guide 4 5 Customer I O Connections...
Страница 68: ...4 8 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...
Страница 80: ...5 12 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...
Страница 85: ...Service and Parts 6 5 Cell Clips Cell Housing Washer Cell O Ring Hex Nut Figure 6 2 Cell replacement...
Страница 87: ...Service and Parts 6 7 Speed Clips Thermocouple Thermocoouple Figure 6 3 Thermocouple replacement...
Страница 91: ...Service and Parts 6 11 Box Heater Block Furnace Wires Heater Figure 6 6b Box Heater Replacement...
Страница 96: ...6 16 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...