4-4 | Thermox WDG-V UOP IECEx/ATEX
Final Daisy Chain Connection Configuration Setting
When connecting multiple analyzers in a daisy chain configuration, en-
able the communication termination resistor (Dip Switch SW-5) on the last
analyzer in the chain (farthest from the user interface). SW-5 is located on
the front of the electronics PCB.
• SW-5 ON: Termination Enabled
• SW-5 OFF: Termination Disabled
The maximum distance between the AMEVision Display User Inter-
face and the analyzer is 4000 feet (1219.2 meters).
Instructions on using the AMEVision software to configure and cali-
brate the analyzer, as well as analyzer status readings, are found in
the separate AMEVision user manual delivered with the analyzer.
Troubleshooting
No Communications:
• Polarity reversed
• Address not set correctly
• Maximum distance exceeded
• Broken connection
Intermittent Communications:
• Poor wiring (i.e., lack of shielded twisted pair)
• Termination enabled on more than one unit (on multi-unit systems)
• Termination not enabled (on multi-unit systems)
• Maximum distance exceeded
Содержание Thermox WDG-V UOP
Страница 34: ...2 6 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...
Страница 65: ...AMEVision Display Installation Guide 4 5 Customer I O Connections...
Страница 68: ...4 8 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...
Страница 80: ...5 12 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...
Страница 85: ...Service and Parts 6 5 Cell Clips Cell Housing Washer Cell O Ring Hex Nut Figure 6 2 Cell replacement...
Страница 87: ...Service and Parts 6 7 Speed Clips Thermocouple Thermocoouple Figure 6 3 Thermocouple replacement...
Страница 91: ...Service and Parts 6 11 Box Heater Block Furnace Wires Heater Figure 6 6b Box Heater Replacement...
Страница 96: ...6 16 Thermox WDG V UOP IECEx ATEX This page intentionally left blank...