Step Settings Display
Adjust
for
Load
Detail
1.
O/P off
Cal count
0V ± 2mV
DVM
O/P off volts
2.
0V
Cal count
0V ± 2mV
DVM
V DAC zero
3.
Max V, 100mA
Cal count
15/30/60V ± 2mV
DVM
V DAC scale
4.
Max V, 100mA
O/P volts
Display = DVM
DVM
V metering scale
5.
I pot at 0
I pot
Display = 0
-
I pot ADC zero
6.
2V, 0mA
Cal count
0mA ± 0.2mA
milli-ammeter
High I DAC zero
7.
2V, 0mA
O/P amps Display = milli-ammeter milli-ammeter
High I metering zero
Change
load
8.
2V, max I
Cal count
1.5/3A ± 1mA
Ammeter
High I DAC scale
9.
2V, max I
O/P amps
Display = ammeter
Ammeter
High I metering scale
10.
2V, 0mA
Cal count
0mA ± 0.02mA
milli-ammeter
Low I DAC zero
11.
2V, 0mA
O/P amps Display = milli-ammeter milli-ammeter
Low I metering zero
12.
2V, 500mA
Cal count
500mA ± 0.1mA
ammeter
Low I DAC scale
13.
2V, 500mA
O/P amps
Display = ammeter
ammeter
Low I metering scale
14.
O/P off
‘End’
-
-
Stores and exits
Maintenance
The Manufacturers or their agents overseas will provide a repair service for any unit developing a
fault. Where owners wish to undertake their own maintenance work, this should only be done by
skilled personnel in conjunction with the service manual which may be purchased directly from
the Manufacturers or their agents overseas.
Cleaning
If the instrument requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent.
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE INSTRUMENT, NEVER
ALLOW WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE NEVER
CLEAN WITH SOLVENTS.