Section 3. Load Connection and Sensing
Remote Sensing
Remote Sensing
WARNING
There is a potential shock hazard at the sense points when using a power supply
with a rated output greater than 40 V. Ensure that connections at the load end are
shielded to prevent contact with hazardous voltages.
CAUTION
Operation of the supply in remote sense mode without the assured connection of
the load wires and remote sense wires to the load may damage the power
supply.
CAUTION
Ground the sense line shield in one place only. Locations include the power
supply’s return output connection at the load, the power supply’s return output at
the negative output terminal, or the power supply’s ground binding post on the
front panel.
Remote sensing permits you to relocate the regulation point of the power supply
from the output terminals to the load or other distribution point terminals.
The power supply provides sense connections beside the output terminals at the front
and rear panels. Use 22-24 AWG twisted, shielded pair wiring to make sense
connections.
With the remote sense leads in place, the supply regulates for the displayed voltage
at the point where the sense lines are connected to the output leads (provided the sum
of these lead losses does not exceed 0.5 V). With the sense lines disconnected, the
supply regulates the voltage at the output terminals. Remote sensing is not required
for constant current mode of operation, where the supply is regulating the output
current.
Note
Do not operate the supply with sense lines connected to the load without also connecting
the normal load power leads to the output terminals.
Always use shielded pair wiring for sense lines to minimize noise effects. (See
“Grounding” on page 17
.)
CAUTION
Do not reverse positive (+) and negative (
) lead connections.
19
Содержание sorensen hpd series
Страница 1: ...HPD Series Programmable DC Power Supply OperationManual TM 30OP 01XN Rev E www programmablepower com...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ii...
Страница 8: ...This page intentionally left blank iv...
Страница 10: ...This page intentionally left blank vi...
Страница 32: ...Section 3 Load Connection and Sensing Remote Sensing 20 Operating Manual for HPD Series Power Supply...
Страница 40: ...APPENDIX RACK MOUNT INSTALLATION INSTRUCTIONS 28...
Страница 41: ...40 29 4X 9 ITEMS 10 AND 13 NOT USED FOR HPD...