Operation
— 22
SPECTROCHECK — 07.05.2015
7.3.2 Measuring
To proceed with a measurement, follow the steps below:
1.
After starting the spectrometer program "Spark Analyzer Pro CHECK", select
the "Analysis" function.
2.
If necessary, change the method by selecting "Load Method" or pressing the
respective function key.
3.
Put the sample in position.
The sample must fully cover the opening in the spark stand plate.
The sample clamp must have contact with the sample (the safety circuit
is closed).
4.
Use "Start" or the respective function key to start the measurement.
The measurement is fully automated.
The measurement is finished automatically.
The measurement results are displayed on the screen.
5.
Perform further measurements if required.
Brush the electrode.
Rotate the sample slightly for each further measurement. The respective
burn spots must not overlap.
6.
Use "Start" or press the respective function key to start the series of
measurement.
The measurement is complete.
7.4 Additional
functions
The following additional functions are available as an option for the
SPECTROCHECK:
Control
samples
Grade
control
Automatic Reproducibility Test
Type
Standardization
Pseudo
Elements
Result
Export
Result
Distribution
For additional information, please contact your responsible sales organization.