User Guide
Page 3 - 7
Genesis WDG-1200/1210
Insitu Oxygen Probes
Mounting Guidelines
1) Mount the control unit to a flat, vertical surface at a height where the
display is clearly visible and the key panel user interface is readily
accessible.
2) There should be a minimum of 150 mm (6 in) clearance below to enable
both the electrical connections and signal cable connections to be made.
3) Ensure that there is an accessible route for connection to the measurement
probe and the remote calibration unit.
4) Maximum length of interconnecting cable between the measurement probe
and the control unit should not exceed 150 meters or 500 feet.
Figure 3-4 Selecting the measurement location