User Guide
Page 3 - 21
Genesis WDG-1200/1210
Insitu Oxygen Probes
Mains Power Supply Plug
NOTE
Ensure there is sufficient clearance to enable connection and disconnection
of the mains connector.
Mains Connector Assembly Procedure
1) Remove approximately 15mm / 0.6 in of the cable’s outer cover to expose
the inner cores.
2) Thread the exposed cores through the cable clamp.
3) Select the correct rubber sealing washer for the cable diameter and thread
the cable through the washer.
4) Thread the cable through the connector body.
5) Connect the individual cores to the screw terminals.
6) Screw the parts of the connector together to ensure that the rubber sealing
washer fits snugly around the cable.
Fitting of Plug Connectors
Do not push or force the plugs onto the connectors. This may result in permanent
damage.
NOTE
The plugs will only fit the connectors in one orientation.
Special lugs on both the plugs and connector should help alignment. The plugs
should attach firmly and secure by twisting the ring.
WARNING
The protective earth conductor terminal on the mains connector
must be properly terminated for operator safety. Continuity of
the protective earth conductor MUST be checked before the unit is
switched on for the first time.
Содержание LAND Genesis WDG-1200
Страница 4: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 6: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 14: ...User Guide Page 2 4 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 68: ...User Guide Page 4 32 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 89: ...User Guide Page 6 9 Genesis WDG 1200 1210 Insitu Oxygen Probes Fig 6 5 Cell and retainer clip placement ...
Страница 94: ......
Страница 95: ......