User Guide
Page 4 - 6
Genesis WDG-1200/1210
Insitu Oxygen Probes
Calibration Screens
Calibration Menu
Ensure that installation of the analyzer is complete (see installation manual) and
that the SYSTEM parameters have been set. If the analyzer is in maintenance
mode the two outer decimal points displayed will flash; during calibration, all four
will flash.
The following information is required:
•
Span gas concentration
•
Zero gas concentration
If the automatic calibration option is selected, the following settings are also
needed:
•
Time until first calibration
•
Time necessary for the calibration gas flow to stabilize
•
Time interval between calibrations
Display Action
CONFIGURATION Sub-menu
Span gas conc.
Enter value
Zero gas conc.
Enter value
Cal time interval
Enter value
Time to next cal
Enter value
Cal settling time
Enter value
Display Action
AUTO CAL. Enter or F1
Display Action
MANUAL CAL.
Zero Enter or F1
Span Enter or F1