User Guide
Page 3 - 15
Genesis WDG-1200/1210
Insitu Oxygen Probes
Cable Recommendations - WDG-1210 Insitu
WDG-1210 Insitu Cable Recommendations
Ref.
Description
Max. o.d.
mm
No. of
Cores
Core
Size mm
2
Core Size
strands/dia
Core
Size awg
Shielded
1
Mains Supply
11
3
0.5
24/0.2
18
No
2
Calibration
13
6
0.2
7/0.2
24
Yes
3
Outputs/Relays
15
17
0.2
7/0.2
24
Yes
4
Probe Power
11
3
0.8
24/0.2
18
Yes
5
Interconnect
13
6
0.5
16/0.2
22
Yes
Cables 1, 3, 4 and 5 are supplied with the analyzer. Cable 2 is supplied with the
remote calibration unit.
Pin Designations - 1200/1210 Insitu
The following is a list of pin designations for Connectors 1, 2, and 3.
Connector Ref. Insert Arrangements
Pin
Function
Wire Colour
1.Mains Power
1
Live
Brown
2
Neutral
Blue
Earth
Green/Yellow
Connector Ref. Insert Arrangements
Pin
Function
2. Remote Calibration Unit
A
H
G
F
E
D
C
B
A
Zero Gas Control
Black
B
Zero Gas Control
Red
C
Span Gas Control
Yellow
D
Span Gas Control
Green
E
Pressure Switch Input +ve Blue
F
Pressure Switch Input 0V
White
G
Not Used
-
H
Not Used
Screen
Содержание LAND Genesis WDG-1200
Страница 4: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 6: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 14: ...User Guide Page 2 4 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 68: ...User Guide Page 4 32 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Страница 89: ...User Guide Page 6 9 Genesis WDG 1200 1210 Insitu Oxygen Probes Fig 6 5 Cell and retainer clip placement ...
Страница 94: ......
Страница 95: ......