1.5 Special Conditions for Safe Use:
Misuse
Should the IPI Mk. II be exposed to overpressure or sudden physical shock
(i.e. being dropped) it should be examined for any damage that may cause
a safety concern. If in doubt please return the unit for evaluation to Ametek.
Please refer to the Customer Service Section for contact information.
WARNING
The IPI Mk. II is not intended for use with flammable substances and is
intended for installation only in locations providing adequate protection
against the entry of solid foreign objects or water capable of impairing
safety.
Caution
To avoid possible damage to calibrator or to equipment under test:
• If the message changes to "OL" the range limit is exceeded and the
pressure source must immediately be removed from the IPI Mk. II to
prevent damage to the pressure transducer inside.
• Maximum torque allowed is 13,5 Nm = 10 ftlbs. NEVER exceed the
torque allowed.