11.0 IPI Mk. II Serial Interface Instructions
Warning
To prevent possible electrical shock, fire, or personal injury,
do not use the RS-232 interface in hazardous areas.
Initiating Communication
The terminal communications can be setup using terminal communication
software on a PC. The terminal settings need to be set as follows:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow control: None
• Local echo on
List of Commands
Command
Description
CAL_START
Puts the calibrator in calibration mode
*CLS
Clears the error queue.
FAULT?
Returns an error code from the error queue
*IDN?
Identification query. Returns the manufacturer, model
number, and firmware revision level of the Calibrator.
TARE
Tares the offset pressure of the reading on the
calibrator
TARE?
Returns the current tare value
PRES_UNIT?
Returns the pressure unit for the upper display.
PRES_UNIT
Sets the pressure unit for the display
ZERO_MEAS
Zeros pressure of the calibrator
ZERO_MEAS?
Returns the current zero offset value
MINMAX_RST
Resets the minimum and maximum recorded values.
MIN?
Returns the minimum recorded value
MAX?
Returns the maximum recorded value
HC_OFF
Turns unit off
HC_DFLT
Sets auto off defaults
TEMP?
Returns temperature in the chosen units
HC_COMP_OFF Turns temperature compensation off.