122770 06
18-09-2007
63
Adjustment
c
Turn the pre-set potentiometer R37 counter clockwise as far
as possible.
d
The calibrator is started in service mode by pressing and
holding
and
while the calibrator is switched on.
e
Press
(SET temp.) and set the calibrator to max.
temperature (250°C, 320°C or 650°C) and press ”ENTER”.
f
Press
(Service) and then
. (Power).
g
R37 is now turned clockwise until DutyK for Primary is
between 0.95 and 0.98.
h
Switch off the calibrator using the power control switch.