20
18-09-2007
122770
06
Error code:
100
Error message:
Calibrator type illegal
Likely cause:
Solution:
Wrong calibrator type
setting
Check that calibrator setting is correct on
Main PCB (see section 2.3.1
on DIP Switch
Setting)
Error code:
200
Error message:
MenuLevel too high
Likely cause:
Solution:
Internal software error
Please report error to AMETEK Denmark
A/S.
Error code:
300
Error message:
AD-converter not responding
Likely cause:
Solution:
Error in communication
with one of the A/D-
converters
Check that all voltages on the power PCB
are correct. If they are, the Main PCB is
defective. Replace the Main PCB.
If not, replace Power PCB.
Error code:
310
Error message:
Slave temperature too high
Likely cause:
Solution:
When main adjustment
has become stable,
differential temperature
between top and bottom
is still outside the limits.
Check measuring circuit with a short circuit
in the differential TC connection (CN16 on
input board) instead of the sensors. Check
in Service Input Menu that Slave TC shows
0°C±0.1°C. If this is OK, there is an error on
the sensor. Otherwise, there is an error on
the Main PCB.
Содержание JOFRA ATC-125
Страница 2: ...2 18 09 2007 122770 06 ...
Страница 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Страница 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Страница 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Страница 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Страница 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...