122770 06
18-09-2007
23
Defective Power PCB
If the 3 LED’s on the Power PCB are not lit,
measure the following voltages on the
terminals of CN3 with the Main PCB
connected.
11 (+) to 12 (-) U > 0,6V
17 (+) to 18 (-): U > 0,6V
19 (+) to 20 (-): U > 0,6V
15 (+) to 16 (-): U > 1,0V
The voltages will appear in pulses.
If these are below the limits change the
Main PCB, adjust and re-calibrate the
calibrator. If not, change and adjust the
Power PCB.
Error:
Calibrator does not cool or heat sufficiently
(ATC-155 only).
Error message:
-
Likely cause:
Solution:
Defective well
Set the SET temperature to –50°C.
On the power PCB measure the voltage
between CN8 terminal 1(+) and terminal 2(-
) and the voltage between CN1 terminal
1(+) and terminal 2(-).
If the voltage on CN1 is higher than 30V
replace the well and re-calibrate the unit.
If the voltage on CN8 is 0.45V ±0.01V and
the calibrator does not cool down
sufficiently, replace the well and re-calibrate
the unit.
Defective Power PCB
If the voltage on CN1 is less than 30V and
the voltage on CN8 differs from 0.45V
±0.01V, adjust R9 until the voltage is 0.45V
±0.01V.
If it is impossible to adjust the voltage on
CN8, replace the Power PCB.
Содержание JOFRA ATC-125
Страница 2: ...2 18 09 2007 122770 06 ...
Страница 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Страница 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Страница 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Страница 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Страница 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...