54
18-09-2007
122770
06
ATC-125
A/B:
In the bottom of the well, unscrew
the 4 nuts holding the sensor cover
and gently remove it.
Remove the outer layer of
insulation - when reassembling a
new insulation layer must be used.
Cut the plastic strap holding
together the wires from the
reference sensor and the 3
thermocouples.
Now the sensor can be retracted.
K.
Replacement of Thermocouples
(Exploded views – fig. 3.2, pos. 21, fig. 3.3 + 3.5 + 3.7,
pos.18)
To replace the thermocoupler in the well unit, follow steps A
through D followed by step F and then follow the procedure
given below:
ATC-156
A/B:
Place the well unit upside down
and remove the 2 screws from the
Connection PCB.
Carefully lift up the Connection
PCB and retract the thermo-
coupler from the well unit.
Содержание JOFRA ATC-125
Страница 2: ...2 18 09 2007 122770 06 ...
Страница 74: ...74 18 09 2007 122770 06 3 3 Exploded View ATC 140 A B 0 35 1 JRL 03 01 2005 A3 ...
Страница 75: ...122770 06 18 09 2007 75 3 4 Exploded View ATC 155 A B ...
Страница 76: ...76 18 09 2007 122770 06 3 5 Exploded View ATC 156 A B ...
Страница 77: ...122770 06 18 09 2007 77 3 6 Exploded View ATC 157 A B ...
Страница 79: ...122770 06 18 09 2007 79 3 8 Exploded View ATC 320 650 A B ...