6-50 | Model 933S UV Analyzer
2. If the analyzer uses
Measured
Pressure Compensation (optional),
change it to
Fixed
mode so that a “
f: Sample gas pressure too low
”
alarm does not occur. To do this, view the
Cell/Misc Parameters
dialog
box and enter a value in the
Default Cell Pres
(pressure compensa-
tion) field that is the same as, or approximately, the Measuring Cell
pressure. Refer to “Cell/Miscellaneous (Compensation) Parameters” in
Chapter 4.
Close the sample inlet valve. Backpurge the analyzer sample system
with Zero gas for 2–3 minutes.
3. Change the
Flow Control
to ‘
4
’ (
shut-in/shut-off
). Click
Apply
(next to
Flow Control
) and then click
OK
.
Bleed down the sample system to atmospheric pressure (check gauge).
Close the valve on the Vent Line (or cap the Vent outlet).
To avoid tearing the Membrane Filters in the Filterblock, always bleed
down the sample system pressure on the downstream (outlet) side of
the Filterblock. Pressure differentials of more than 210 KPA (30 PSI)
from the inlet side to the outlet side can tear the Filters and they will
have to be replaced.
4. Disconnect AC power to the analyzer and power down the computer
before performing maintenance on the analyzer.
5. Remove the Optical Bench from the Upper Enclosure:
While removing the Optical Bench from the Upper Enclosure, grasp
it by the upper portion of the Optical Bench to support it; do not sup-
port the assembly by its circuit boards or source lamps. Also, take care
not to damage the enclosure’s joining areas (flamepath).
a. Remove the (24) M10 screws from the Upper Enclosure door and
open it.
b. Disconnect the following wiring connectors from the
Micro-Inter-
face
board (see Figure 6-12 for connector locations):
• Measuring Cell RTD line from
J300
.
• RS-422 communication line from
J104
.
• Micro-Interface AC power line from
J200
.
• DC power line from
J103
.
•
Optional:
If your analyzer has a Cell Heater, disconnect the
Heater Cartridge wire (marked ‘
N
’) at
J200
(Terminal 2) and
the Overtemp Switch wire (marked ‘
L
’) at
J200
(Terminal 1).
!
CAUTION
!
CAUTION
Setup (tab)
Cell/Misc
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...