Controller / User Interface
|
4-41
Sensor
Enabled
Check this to enable the sensor assigned to this circuit (example,
for CO
2
measurement). All parameters and action buttons for a
sensor are ignored if the sensor is not enabled.
Species Name
Enter the species being measured by the sensor. This will be dis-
played in other software screens.
Zero-Cal Signal Setpoint
Target
The level, in mV, to which the sensor Zero calibration attempts to
adjust the sensor reference signal toward. For Sensor1, this param-
eter is only applicable to “Differential” Signal Type (it is ignored for
Sensor1 with “Voltage” Signal Type).
Delta (+\-)
(
Sensor1
and
Sensor3
screens)
The deviation, in mV, from the target level for which the sensor
Zero calibration uses to determine successful completion. For Sen-
sor1, this parameter is only applicable to “Differential” Signal Type
(it is ignored for Sensor1 with “Voltage” Signal Type). This value
should be between 50–250 after the Zero.
Signal Level at Zero-Cal
Reference
(
Sensor1
screen with
Differential Signal Type
selected and
Sensor3
screen)
The sensor
Reference
channel signal level, in mV, at
the completion of the Zero calibration. Record this value after the
Zero.
Differential
(
Sensor1
and
Sensor3
screens with
Differential
Signal
Type
selected) The sensor
Differential
channel signal level, in mV,
at the completion of the Zero calibration. Record this value after
the Zero.
Voltage
(
Sensor1
and
Sensor3
screens with
Voltage
Signal Type
se-
lected
and
Sensor2
screen) The reading (in mV) recorded by the
analyzer at the completion of the Zero calibration. This parameter
is automatically updated.
Pressure
(
Sensor2
screen and
Sensor3
screen with
Voltage
Signal Type
selected) The Measuring Cell (Sample gas) pressure (in mmHg)
recorded by the analyzer at the completion of the Zero calibration.
This parameter is automatically updated. If the
Pressure Correc-
tion Source
for the sensor is
Default
, the
Default Cell Pres
is used.
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...